10-50 ADS Manual
Peak Velocity Subsystem
The following tables contain general troubleshooting techniques for
the velocity subsystem corresponding to the Peak Combo Sensor
and Surface Combo Sensor (
Surcharge Velocity Sensor only
).
Since the Surface Combo Sensor is mounted at the top of the pipe,
potential issues involving a sensor that is out of the flow do not
apply to the Surface Combo Sensor.
Note:
Contact your regional ADS representative for
further diagnosis prior to replacing a sensor.
Problem
Monitor often provides a velocity reading of 0.
Possible
Causes
Sensor’s velocity transmit/receive surfaces may be dirty
or the velocity function is broken.
Minimum flows may be occurring within the sensor
deadband (extending 1.0 inches (25 mm) up from the
bottom of the Peak Combo Sensor).
Cabling between the processor board and the velocity
sensor may be bad.
Velocity parameters may require adjustment.
Possible
Solutions
Clean the sensor’s transmit/receive surfaces.
Verify that minimum flows are occurring outside the
sensor deadband.
Check the velocity parameters.
Inspect the sensor cables for tightness.
Check for moisture in the connector.
Contact your regional ADS representative.
Replace the sensor.
Replace the monitor if defective.
Содержание Intrinsically-Safe TRITON+
Страница 19: ...1 12 ADS TRITON Manual Control Drawing ...
Страница 25: ...1 18 ADS TRITON Manual First page of the TRITON ATEX Certificate ...
Страница 27: ...1 20 ADS TRITON Manual First page of the TRITON IECEx Certificate of Conformity ...
Страница 29: ...1 22 ADS TRITON Manual CSA Certificate of Compliance Page 1 ...
Страница 30: ...Introduction 1 23 CSA Certificate of Compliance Page 2 ...
Страница 32: ...Introduction 1 25 Sample TRITON Declaration of Conformity Page 2 ...
Страница 34: ...Introduction 1 27 Activate Monitor Qstart software Run Diagnostics and Perform Confirmations Qstart software ...
Страница 49: ...2 12 ADS TRITON Manual ExPAC XBUS ...
Страница 159: ...External Power 5 9 Power supply firmly secured to the DIN rail Front view of power supply mounted to DIN rail ...
Страница 171: ...External Power 5 21 POWER SUPPLY DC OUTPUT TERMINAL CONNECTIONS Wires exiting DC output terminal on power supply ...
Страница 231: ...6 44 ADS TRITON Manual Editing the Smart Ultrasonic Depth Sensor Device Edit Smart Depth 1 Parameters dialog ...