Configuration and Activation 6-77
The monitor takes the sensor measurements and displays the
readings in the Results window. The results may include the
specific reading (e.g., depth or velocity), the quality of the
reading, the maximum signal strength, and the temperature.
When taking depth measurements, the dialog also displays a
cross section of the pipe, offering a visual perspective of the
depth of the flow.
Note:
Pressure depth readings compensate for the
physical and electronic offsets, and the
Automatic Pressure
Calibration
value (PRESSK and PRESSK2).
8.
Verify the accuracy, consistency, and quality of the readings,
and then edit the corresponding device parameters as necessary.
9.
(
optional
) Repeat steps 3 through 10 for every other sensor on
the device dialog on which you want to run diagnostics.
10.
Click on the
OK
button once you are finished running
diagnostics for all the appropriate sensors corresponding to the
current device. This saves any changes made to the parameters
to the local directory or network. Clicking on the
Cancel
button closes the device dialog without saving any parameter
modifications.
11.
If any changes have been made to
any
of the device parameters,
you
must
reactivate the monitor by clicking on the
Activate
button in the
Functions
section.
Note:
There is no need to reactivate a monitor if there
were no parameter changes!
12.
Once activation is complete, the status bar displays
Activate
successful
.
13.
Click on the
Hangup
button in the
Functions
section when you
are finished running diagnostics on sensors associated with the
current monitor.
14.
Click on the
Save
button to save any changes in the device
parameters to the local directory or network.
Содержание Intrinsically-Safe TRITON+
Страница 19: ...1 12 ADS TRITON Manual Control Drawing ...
Страница 25: ...1 18 ADS TRITON Manual First page of the TRITON ATEX Certificate ...
Страница 27: ...1 20 ADS TRITON Manual First page of the TRITON IECEx Certificate of Conformity ...
Страница 29: ...1 22 ADS TRITON Manual CSA Certificate of Compliance Page 1 ...
Страница 30: ...Introduction 1 23 CSA Certificate of Compliance Page 2 ...
Страница 32: ...Introduction 1 25 Sample TRITON Declaration of Conformity Page 2 ...
Страница 34: ...Introduction 1 27 Activate Monitor Qstart software Run Diagnostics and Perform Confirmations Qstart software ...
Страница 49: ...2 12 ADS TRITON Manual ExPAC XBUS ...
Страница 159: ...External Power 5 9 Power supply firmly secured to the DIN rail Front view of power supply mounted to DIN rail ...
Страница 171: ...External Power 5 21 POWER SUPPLY DC OUTPUT TERMINAL CONNECTIONS Wires exiting DC output terminal on power supply ...
Страница 231: ...6 44 ADS TRITON Manual Editing the Smart Ultrasonic Depth Sensor Device Edit Smart Depth 1 Parameters dialog ...