2 Safety | 2.5 Safety Devices and Interlocks
2.5.1 Protective Cover Panels
Due to hazardous voltages and X-rays inside the microscope, the microscope is equipped with
protective cover panels.
3
1
2
Fig. 1: Protective cover panels
1
Electron optical column protective cover
panels
2
Specimen chamber protective cover
panels
3
Plinth protective cover panels
Operation of the microscope is only allowed with attached protective cover panels.
2.5.2 Main Disconnect Device
To disconnect the microscope with all its components from the mains supply, unplug the CEE con-
nector (blue for 208–230 VAC).
Fig. 2: CEE connector
If the Emergency Off (EMO) Option is installed, the Main Switch of the EMO Option may be used
as Main Disconnect Device for the microscope and its components, refer to
Instruction Manual ZEISS SIGMA series | en-US | Rev. 7 | 352102-9344-006