2 Safety | 2.4 Warning Labels and Lights
ZEISS
The safety labels always need to be attached to the correct spots at the microscope. If a safety la-
bel is lost or unreadable, it needs to be reordered via the following reorder numbers:
Position and Figure of the Safety La-
bel
Description
1
At the front of the chamber door.
Only present if the optional Raman Spectrometer
is installed.
Laser Radiation
Class 1M laser product
Do not view directly with optical instruments.
Diffuse laser reflection possible when chamber
door is open.
Reorder no. 344700-0033-000-15en
2
At the front of the chamber door.
347800-0007-000
Nitrogen Hazard
WARNING
Danger of suffocation
Ensure area around
instrument is sufficiently
ventilated
Nitrogen Hazard
Danger of suffocation
Ensure area around instrument is sufficiently ven-
tilated
Reorder no. 347800-0007-000
3
At the front of the chamber door.
CAUTION
Risk of injury
347800-0033-000-02en
Fingers could be trapped.
Always close the chamber door
before you move the stage.
Risk of injury
Fingers could be trapped. Always close the cham-
ber door before you move the stage.
Reorder no. 347800-0033-000-02en
4
At the front of the chamber door.
Risk of damage
CAUTION
347800-0033-000-04en
The FESEM or specimen could be
damaged if specimen stage is at a
short working distance.
Move the specimen stage to a long
working distance before opening
the chamber door.
Risk of damage
The FESEM or specimen could be damaged if
specimen stage is at a short working distance.
Move the specimen stage to a long working dis-
tance before opening the chamber door.
Reorder no. 347800-0033-000-04en
5
At the front of the plinth
Avoid injury
CAUTION
347800-0033-000-01en
Make sure you have read and
understood the instruction
manual before operating this
product.
Avoid injury
Make sure you have read and understood the in-
struction manual before operating this product.
Reorder no. 347800-0033-000-01en
6
At the rear of the microscope
347800-0018-000
Interaction with metallic objects
may produce Pinch Hazards.
Persons with Medical Implants
KEEP BACK 12 inches.
Magnetic Field
CAUTION
Magnetic Field
Interaction with metallic objects may produce
Pinch Hazards.
Persons with Medical Implants KEEP BACK 12
inches.
Reorder no. 347800-0018-000