ZEISS
5 Commissioning and First Operating Steps | 5.8 Working with Optional Accessories
5. Use the buttons in the
Manual stage control
section of the
SEM Raman Correlation
win-
dow to refine the adjustment.
INFO:
If you use the
Manual stage control
,
then the backlash correction is automatically
activated. The backlash correction ensures the
required relative accuracy of the stage.
à
In the
Raman Correlation
tab, a message is
displayed that asks you to focus the image.
7. Focus the image.
8. In the
Raman Correlation
tab of the
SEM Ra-
man Correlation
window, click
Move to Ra-
man
The following steps are executed automatically:
à
The image of the USB TV camera is automati-
cally displayed to let you see the movement
towards the Raman objective and the Raman
working distance.
INFO:
The speed of the dual joystick de-
pends on the magnification. If the image of
the USB TV camera is displayed, then the
stage travels with maximum speed.
à
The stage moves the specimen to the Raman position.
à
After the movement is finished, the SEM image is displayed again.
INFO:
Not all stage positions are within the reach of both, the SEM and the Raman mi-
croscope. In this case, a message is displayed in the message field. Often a rotation of
the specimen towards the Raman microscope helps. Use the compucentric rotation to ro-
tate the specimen.
9. Check that the objective is in the position x = 0, y = 0, and z = 0.