App-7
IM DL850-03EN
Appendix
1
2
3
4
5
6
App
Index
(Continued from previous page)
Record Length
4 Gpoint
5 Gpoint
10 Gpoint
20 Gpoint
50 Gpoint
Time/div Sample
Rate
(S/s)
Display
Record
Length
(Points)
Sample
Rate
(S/s)
Display
Record
Length
(Points)
Sample
Rate
(S/s)
Display
Record
Length
(Points)
Sample
Rate
(S/s)
Display
Record
Length
(Points)
Sample
Rate
(S/s)
Display
Record
Length
(Points)
6 h
10 k
2.16 G
20 k
4.32 G
20 k
4.32 G
50 k
10.8 G
200 k
43.2 G
7 h
10 k
2.52 G
10 k
2.52 G
20 k
5.04 G
50 k
12.6 G
100 k
25.2 G
8 h
10 k
2.88 G
10 k
2.88 G
20 k
5.76 G
50 k
14.4 G
100 k
28.8 G
9 h
10 k
3.24 G
10 k
3.24 G
20 k
6.48 G
50 k
16.2 G
100 k
32.4 G
10 h
10 k
3.6 G
10 k
3.6 G
20 k
7.2 G
50 k
18 G
100 k
36 G
12 h
5 k
2.16 G
10 k
4.32 G
20 k
8.64 G
20 k
8.64 G
100 k
43.2 G
1 day
2 k
1.728 G
5 k
4.32 G
10 k
8.64 G
20 k
17.28 G
50 k
43.2 G
2 days
2 k
3.456 G
2 k
3.456 G
5 k
8.64 G
10 k
17.28 G
20 k
34.56 G
3 days
1 k
2.592 G
1 k
2.592 G
2 k
5.184 G
5 k
12.96 G
10 k
25.9 G
When the time axis setting is 100 ms or greater (the settings surrounded by bold lines) and the trigger mode is Auto or Auto Level,
waveforms are displayed in roll mode.
Appendix 1 Relationship between the Time Axis Setting, Record Length, and Sample Rate