6-28
IM DL850-03EN
Item
Specifications
Temperature coefficient
[Voltage measurement]
Zero point:
Gain:
±(0.01% of 10 div)/°C (Typical
6
)
±(0.02% of 10 div)/°C (Typical
6
)
Reference junction compensation
accuracy (when the input terminal
temperature is balanced)
K, E, J, T, L, U, N:
R, S, B, W:
Au7Fe:
±1°C
±1.5°C
±1K
Bandwidth limit
[Temperature measurement]
[Voltage measurement]
(Digital analog filter)
Select from OFF, 30 Hz, 8 Hz, and 2 Hz + 150 Hz secondary
analog filter
Select from OFF, AUTO, 4 kHz, 400 Hz, or 40 Hz.
Cutoff characteristics: −12 dB/OCT (Typical,
6
setting other than
AUTO)
Cutoff frequency (fc) when set to AUTO (701262 only)
Sample Rate
Cutoff Frequency (fc)
100 kS/s or higher 40 kHz
100 S/s to 50 kS/s
40% of the sample rate
50 S/s or less
20 Hz
Cutoff characteristics for AUTO: −65 dB at 2 × fc (Typical
6
)
Table of cutoff frequency
characteristics of the
anti-aliasing filter (AAF)
When the filter is set to Auto, the anti-aliasing filter and low-pass filter are automatically set
according to the sample rate.
Sample Rate
AAF
Low-Pass Filter
100 kS/s
40 kHz
OFF
50 kS/s
20 kHz
OFF
20 kS/s
8 kHz
OFF
10 kS/s
4 kHz
4 kHz
5 kS/s
2 kHz
4 kHz
2 kS/s
800 Hz
4 kHz
1 kS/s
400 Hz
400 Hz
500 S/s
200 Hz
400 Hz
200 S/s
80 Hz
400 Hz
100 S/s
40 Hz
40 Hz
50 S/s
20 Hz
40 Hz
20 S/s to 5 S/s
20 Hz
40 Hz
2 S/s or less
20 Hz
40 Hz
Ext sample
40 kHz
OFF
1 Value measured under standard operating conditions (section 19.11).
2 Does not include the reference junction temperature compensation accuracy.
3 This module supports Au7Fe with 0.07% metal content with respect to gold.
H
L
5
4
6. Typical value represents a typical or average value. It is not strictly warranted.
WARNING
• Do not apply input voltage exceeding the maximum input voltage or allowable common
mode input voltage.
• To prevent the possibility of electric shock, be sure to furnish protective earth grounding of
the DL850/DL850V.
• To prevent the possibility of electric shock, be sure to fasten the module screws.
Otherwise, the electrical and mechanical protection functions will not be activated.
6.13 Module Specifications