6-15
IM DL850-03EN
Specifications
1
2
3
4
5
6
App
Index
6.12 General Specifications
Item
Specifications
Standard operating
conditions
Ambient Temperature:
23 ± 5°C
Ambient humidity:
20 to 80% RH
Supply voltage and
frequency errors:
Within ±1% of rating
After the DL850/DL850V has been warmed up for 30 minutes and then calibration has been
performed
Recommended calibration period 1 year
Warm-up time
At least 30 minutes
Storage environment
Temperature:
−20 to 60°C
Humidity:
20 to 85% RH (no condensation)
Altitude:
3000 m or less
Operating environment
Temperature:
5 to 40°C
Humidity:
20 to 85% RH (when the printer is not being used)
35 to 85% RH (when the printer is used)
In both cases, no condensation
Altitude:
2000 m or less
Standard model
Rated supply voltage
100 to 120 VAC, 220 to 240 VAC (auto switching)
Permitted supply voltage
range
90 to 132 VAC, 198 to 264 VAC
Rated supply frequency 50/60 Hz
Permitted supply voltage
frequency range
48 to 63 Hz
Power fuse
Built in (not replaceable)
Maximum power
consumption
200 VA
Withstand voltage
1500 VAC for one minute between the power supply and case
Insulation resistance
10 MΩ or more for 500 VDC between the power supply and case
External dimensions
Approx. 355 mm (W) × 259 mm (H) × 180 mm (D), not including the handle and protrusions
Weight
Approx. 6.5 kg (weight of the DL850/DL850V only without paper and with the /M2, /HD1, /C1, and
/P4 options installed)
Instrument cooling
method
Forced air cooling. Exhaust on the left side and top panel.
DL850V DC power supply model (/DC option; the AC power supply specifications are the same as those shown for the standard
model)
Supply format
Auto DC/AC switching (AC preferred), isolation between DC power input terminal and the DL850V
Rated supply voltage
12 VDC
Permitted supply voltage 10 to 18 VDC
Power consumption
Approx. 90 VA (typical
1
):
When driving 2 channels (High-Speed 10 MS/s, 12-Bit Isolation
Module × 1)
Approx. 105 VA (typical
1
):
When driving 8 channels (High-Speed 100 MS/s, 12-Bit Isolation
Module × 2 and High-Speed 10 MS/s 12-Bit Isolation Module × 2)
Approx. 135 VA (typical
1
):
When driving 16 channels (High-Speed 100 MS/s, 12-Bit Isolation
Module × 4 and High-Speed 10 MS/s 12-Bit Isolation Module × 4)
Maximum power
consumption
Approx. 150 VA
Standby power
(during DC standby)
30 mW (typical
1
, DC power consumption when power is supplied both to AC and DC power)
Voltage input protection
circuit
Overcurrent detection: Breaker (15 A)
Reverse connection protection: Breaker shutdown
Undervoltage detection: Cut off at a voltage less than approx. 9.5 V
Overvoltage detection: Cut off at a voltage greater than approx. 18 V
Withstand voltage
30 VDC for one minute between the DC power input terminal and earth terminal
External dimensions
Approx. 355 mm (W) × 250 mm (H) × 200 mm (D),
(DL850V with the /DC option installed, not including the handle and protrusions)
Weight
Approx. 7.4 kg (only the DL850V with all options (/M3/C8/C10/P4/DC options)
Approx. 9.8 kg (DL750 (with /DC option) + eight High-speed 10 MS/s, 12-Bit Isolation Modules)
Weight increase by the DC power option Approx. 800 g
Indicator function
Indicates the status of the power supply to the DL850V using two LEDs
Instrument cooling
method
Forced air cooling. Exhaust on the left side and top panel.
In DC mode, air is vented from the left side, top panel, and rear panel.