App-1
IM DL850-03EN
Appendix
1
2
3
4
5
6
App
Index
Appendix 1 Relationship between the Time Axis
Setting, Record Length, and Sample Rate
When the High-Speed 100 MS/s, 12-Bit Isolation Module Is Installed
When the Record Length Is 1 kpoint, 2.5 kpoint, 5 kpoint, 10 kpoint, or 25 kpoint
Record
1 kpoint
2.5 kpoint
5 kpoint
10 kpoint
25 kpoint
Time/div Sample
Rate
(S/s)
Display
Record
Length
(Points)
Sample
Rate
(S/s)
Display
Record
Length
(Points)
Sample
Rate
(S/s)
Display
Record
Length
(Points)
Sample
Rate
(S/s)
Display
Record
Length
(Points)
Sample
Rate
(S/s)
Display
Record
Length
(Points
)
100 ns
100 M
100
100 M
100
100 M
100
100 M
100
100 M
100
200 ns
100 M
200
100 M
200
100 M
200
100 M
200
100 M
200
500 ns
100 M
500
100 M
500
100 M
500
100 M
500
100 M
500
1 μs
100 M
1 k
100 M
1 k
100 M
1 k
100 M
1 k
100 M
1 k
2 μs
50 M
1 k
100 M
2 k
100 M
2 k
100 M
2 k
100 M
2 k
5 μs
20 M
1 k
50 M
2.5 k
100 M
5 k
100 M
5 k
100 M
5 k
10 μs
10 M
1 k
20 M
2 k
50 M
5 k
100 M
10 k
100 M
10 k
20 μs
5 M
1 k
10 M
2 k
20 M
4 k
50 M
10 k
100 M
20 k
50 μs
2 M
1 k
5 M
2.5 k
10 M
5 k
20 M
10 k
50 M
25 k
100 μs
1 M
1 k
2 M
2 k
5 M
5 k
10 M
10 k
20 M
20 k
200 μs
500 k
1 k
1 M
2 k
2 M
4 k
5 M
10 k
10 M
20 k
500 μs
200 k
1 k
500 k
2.5 k
1 M
5 k
2 M
10 k
5 M
25 k
1 ms
100 k
1 k
200 k
2 k
500 k
5 k
1 M
10 k
2 M
20 k
2 ms
50 k
1 k
100 k
2 k
200 k
4 k
500 k
10 k
1 M
20 k
5 ms
20 k
1 k
50 k
2.5 k
100 k
5 k
200 k
10 k
500 k
25 k
10 ms
10 k
1 k
20 k
2 k
50 k
5 k
100 k
10 k
200 k
20 k
20 ms
5 k
1 k
10 k
2 k
20 k
4 k
50 k
10 k
100 k
20 k
50 ms
2 k
1 k
5 k
2.5 k
10 k
5 k
20 k
10 k
50 k
25 k
100 ms
1 k
1 k
2 k
2 k
5 k
5 k
10 k
10 k
20 k
20 k
200 ms
500
1 k
1 k
2 k
2 k
4 k
5 k
10 k
10 k
20 k
500 ms
200
1 k
500
2.5 k
1 k
5 k
2 k
10 k
5 k
25 k
1 s
100
1 k
200
2 k
500
5 k
1 k
10 k
2 k
20 k
2 s
50
1 k
100
2 k
200
4 k
500
10 k
1 k
20 k
3 s
20
600
50
1.5 k
100
3 k
200
6 k
500
15 k
4 s
20
800
50
2 k
100
4 k
200
8 k
500
20 k
5 s
20
1 k
50
2.5 k
100
5 k
200
10 k
500
25 k
6 s
10
600
20
1.2 k
50
3 k
100
6 k
200
12 k
8 s
10
800
20
1.6 k
50
4 k
100
8 k
200
16 k
10 s
10
1 k
20
2 k
50
5 k
100
10 k
200
20 k
20 s
5
1 k
10
2 k
20
4 k
50
10 k
100
20 k
30 s
5
1.5 k
10
3 k
20
6 k
50
15 k
1 min
5
3 k
10
6 k
20
12 k
2 min
5
6 k
20
24 k
3 min
5
9 k
10
18 k
4 min
10
24 k
5 min
5
15 k
6 min
5
18 k
7 min
5
21 k
When the time axis setting is 100 ms or greater (the settings surrounded by bold lines) and the trigger mode is Auto or Auto Level,
waveforms are displayed in roll mode.
Appendix