6-9
IM DL850-03EN
Specifications
1
2
3
4
5
6
App
Index
Item
Specifications
LPF:
Cutoff frequency: 3 MHz to 0.2 Hz
Resolution:
3 MHz to 300 kHz, in steps of 20 kHz
Resolution:
298 kHz to 30 kHz, in steps of 2 kHz
Resolution:
29.8 kHz to 3 kHz, in steps of 200 Hz
Resolution:
2.98 kHz to 0.3 kHz, in steps of 20 Hz
Resolution:
298 Hz to 30 Hz, in steps of 2 Hz
Resolution:
29.8 Hz to 0.2 Hz, in steps of 0.2 Hz
HPF:
Cutoff frequency: 3 MHz to 20 Hz
Resolution:
3 MHz to 300 kHz, in steps of 20 kHz
Resolution:
298 kHz to 30 kHz, in steps of 2 kHz
Resolution:
29.8 kHz to 3 kHz, in steps of 200 Hz
Resolution:
2.98 kHz to 20 Hz, in steps of 20 Hz
BPF:
Center frequency: 3 MHz to 60 Hz
Resolution:
3 MHz to 120 kHz, in steps of 20 kHz
Resolution:
118 kHz to 12 kHz, in steps of 2 kHz
Resolution:
11.8 kHz to 1.2 kHz, in steps of 200 Hz
Resolution:
1.18 kHz to 60 Hz, in steps of 20 Hz
Bandwidth:
2, 1.5, 1 MHz, 500, 200, 150, 100, 50, 20, 15,
10, 5, 2, 1.5, 1 kHz, 500, 200, 100 Hz (there
are limits based on the center frequency)
Cutoff characteristic: –24 dB/Oct
Phase:
Nonlinear phase characteristicst
Interpolation:
Interpolation can be turned on and off.
Resolver:
The angle of rotation is computed from the carrier signal, sine signal, and cosine
signal.
Supported carrier frequencies: 1 kHz to 20 kHz
Tracking filter band setting: 2 kHz, 1 kHz, 250 Hz, 100 Hz
Maximum measurable angular acceleration:
140000 rps
2
(when the tracking filter is set to 2 kHz)
54000 rps
2
(when the tracking filter is set to 1 kHz)
1800 rps
2
(when the tracking filter is set to 250 Hz)
180 rps
2
(when the tracking filter is set to 100 Hz)
Demodulation of the pulse width modulated (PWM) signal:
The PWM-modulated signal is integrated over the modulation period and converted
into an analog signal.
Reactive power
The reactive power is computed from apparent power and effective power.
CAN ID detection
The frame of the CAN bus signal with the specified ID is detected.
Math source
waveforms
• Analog channels 1 to 16 and the sub channels of scanner and CAN modules.
• Math results can be specified as sources of another channel.
• However, you can only specify math results of channels whose numbers are smaller
than the channel that you are specifying sources for.
Math delay
A uniform delay for each math operation, regardless of the number of math channels
Mean
The mean can be calculated on the math results.
Number of taps: 32 (fixed)
The calculation period changes automatically according to the DL850/DL850V sample
rate.
Zoom Method
Zooming by
Setting a
Magnification
V Zoom is set in steps of 1, 2, or 5 to a value in the range of 10.00E–21 to 500.0E+18.
Zooming by
Setting Upper and
Lower Display
Limits
Upper limit:
±5.0000E+22; lower limit: ±1.0000E–23
6.5 Features