Initializing the Micro-Tech
Initializing the Software
Thermo Fisher Scientific
Micro-Tech 9101/9201 Reference Manual, Rev J
2-37
6.
Press the Enter button and the following screen appears.
7.
Press the Edit button. Use the keypad to enter the number you
calculated earlier (and entered in
) for the
“Number of revolutions for test.” (We entered 5 revolutions, as
described in our example on page 2-11.)
8.
Press the Enter button and the following screen appears. (The default
value is 30 seconds.)
9.
Press the Edit button. Use the keypad to enter the number you
calculated earlier (and entered in
) for the
BATCH
Enter the number
of belt revolutions
to be timed 1 rev
READY
ALARM
CALIB
ABORT
EDIT
BATCH
Enter time for
revolutions to pass
reference 30 sec
READY
ALARM
CALIB
ABORT
EDIT