C-68
Micro-Tech 9101/9201 Reference Manual, Rev J
Thermo Fisher Scientific
If the question is not answered, the screen disappears after ten seconds,
and the Micro-Tech assumes the answer is NO. HW (hardware) alarm is
on and cannot be reset. The screen will appear each time power is cycled
if the question is not answered.
Answer YES because this is a hardware configuration change. Set-up data
must now be entered.
Through the new PROFIB menu, which can be found on Main Menu 5, it
is possible to assign the slave address:
The limits of values that can be set are the following:
Default:
1
Min:
1
Max:
126
(Password: Service.)
In the next two scrolls, the operator can define independently the read and
write buffer dimensions.
The limits of values that can be set are the following:
Slave address
Buffer Dimension
BATCH
- PROFIBUS SCROLL 2 -
Read Buffer dim.
48
READY
ALARM
CALIB
EDIT
BATCH
- PROFIBUS SCROLL 1 -
Address
1
READY
ALARM
CALIB
EDIT