background image

 

C-74

     Micro-Tech 9101/9201 Reference Manual, Rev J

 

Thermo Fisher Scientific 

The master system has the possibility to detect a communication error 
condition by checking the diagnostics.  In case of an error, the Micro-
Tech will activate the request for diagnostic acquisition, indicating the 
expanded device related diagnostics data is present. 

When the master requires the diagnostics, it receives: 

 

Table C–7.

 Composition of the Diagnostics Data 

Byte 

Bit Position 

Diagnostic Data 

 

 

 

 

 

 

 

 

StatStatus1 

 

 

 

 

 

 

 

 

StatStatus2 

 

 

 

 

 

 

 

 

StatStatus3 

 

 

 

 

 

 

 

 

Master Add 

 

 

 

 

 

 

 

 

Ident NumberHigh 

 

 

 

 

 

 

 

 

Ident NumberLow 

Ext Diag - Header 

GE 

RE 

LE 

CE 

Ext Diag - Data 

 

GE  Group Error: 

Invalid group identifier, the group does not 
exists 

RE  Register number Error: 

The number of requested register is    

 

 

 

wrong. 

LE  Limits Error: 

A write operation has been performed but the 
value to write overlaps the limits. 

CE  Coherency Error: 

The master has tried to write or read partially 
a variable composed by more registers (e.g., 
only one word of a floating variable). 

 

The individual bits i

Table C–7

 

have the following meaning: 

Bit 7: 

Diag.Master_Lock

 

The DP-Slave has been parameterized from another 
master.  This bit is set by the DP-Master (class 1), if the 
address in octet 4 is different from 255 and different from 
the own address.  The DP-Slave sets this bit to zero. 

 

Bit 6: 

Diag.Prm_Fault

 

Diagnostic Data 

Summary of Contents for Micro-Tech 9101

Page 1: ...Micro Tech 9101 9201 Integrator Reference Manual REC 4281 Rev J Part Number 127336 English...

Page 2: ...ONLY for motherboard pulse output Rev E May 2013 3322 Corrected error on motherboard jumpers Dual plant A D jumpers Removed 9105 information Added 9201 information Rev F September 2013 3363 Correctio...

Page 3: ...gram only in the country where you acquired the equipment as defined below You obtain no rights other than those granted you under this license The term equipment means the equipment with which the so...

Page 4: ...license is as follows Thermo Fisher Scientific will a replace defective media or b make a warranted software program operate or c replace the software program with a functionally equivalent software...

Page 5: ...PERSONAL SAFETY CARE MUST BE TAKEN WHEN WORKING ON OR AROUND THE MICRO TECH AS WITH ALL SUCH DEVICES THE MAIN SUPPLIES ELECTRICAL AND OTHER TO THE SYSTEM MUST BE LOCKED OFF WHEN PERFORMING REPAIR OR M...

Page 6: ...and corrective maintenance which normally occur in the operation of the equipment EXCEPT FOR THOSE WARRANTIES SPECIFICALLY CONTAINED HEREIN SELLER DISCLAIMS ANY AND ALL WARRANTIES WITH RESPECT TO THE...

Page 7: ...o Tech screens themselves please consult the glossary at the end of the manual Conventions The following conventions are used in this manual The names of Micro Tech buttons functions and so on are sho...

Page 8: ...sages for your Micro Tech and its associated scale system Please read all safety messages very carefully because this information is important for your own personal safety and the safety of others WAR...

Page 9: ...Connecting the Incoming Power Supply 1 10 Configuring Jumpers and Switches 1 11 Micro Tech Features 1 11 Standard Features 1 11 Inputs and Outputs 1 12 Micro Tech Menus and Functions 1 12 Measuring F...

Page 10: ...mple 2 11 Manually Determine Test Duration 2 12 The Next Steps 2 13 Using the Console 2 13 Initializing the Software 2 13 Overview 2 13 Cold Starting the Micro Tech 2 14 Setting the Date 2 14 Setting...

Page 11: ...Soft Key Buttons 3 2 Status LEDs 3 2 Running the Micro Tech 3 2 Run Screen 3 3 Viewing the Totals 3 3 Resetting the Totals 3 4 Resetting the Reset and Operator Totals 3 4 Calibrating the Micro Tech 3...

Page 12: ...A 8 Partial Belt Length Method A 8 Full Belt Length Method A 10 Material Factoring A 12 Reset Weight Factor A 13 Static Weight A 14 Auto Span A 15 Record Results A 16 Motherboard Terminal Block Defin...

Page 13: ...nu B 22 Main Menu 3 B 25 Protection Menu B 25 Decreasing the Protection Level B 26 On Line Procedure for Changing Protection Level B 27 Protection Menu Table B 28 Diagnostics Menu B 29 Test Menu B 30...

Page 14: ...ption C 28 Register Mapping C 32 Allen Bradley DF1 C 35 Description C 35 Register Mapping C 41 EtherNet Port C 44 Ethernet IP C 44 IP Address C 45 Controller Tag Listing Example C 45 Ladder Diagram Ex...

Page 15: ...rence Manual Rev J ix Data Organization C 66 Read Operations C 67 Write Operations C 68 Header C 68 Data C 69 Diagnostic Data C 70 Register Mapping C 72 Block 0 C 72 Block 1 C 72 Block 2 C 73 Block 10...

Page 16: ...0 Location of Load Cell Data 2 10 Figure 2 11 Belt Scale Code Entry Screen 2 28 Figure 3 1 Main Features of the Micro Tech Console 3 1 Figure 3 2 Run Screen 3 3 Figure 4 1 Four Wire Load Cell No Sense...

Page 17: ...r Scientific Micro Tech 9101 9201 Reference Manual Rev J xi Figure C 1 Thermo Fisher Static Home Page C 44 Figure C 2 Thermo Fisher Test Page Example C 45 Figure C 3 Modbus Screen C 58 Figure C 4 Prof...

Page 18: ...ble A 3 Premium A D Board Load Cell Technical Specifications Model 9201 A 28 Table C 1 Type Limits and Format of Registers C 7 Table C 2 Maximum Cable Length of a Segment C 61 Table C 3 Bus Connector...

Page 19: ...an be safely stored with its cover and latches secured and with the hole plugs installed During storage do not expose the Micro Tech to moisture or to temperatures outside the range of 22 to 158 F 30...

Page 20: ...a Fault output In addition many automatic and check functions are available to monitor its calibration functions and maintenance schedule There are two models of Micro Tech the field mounted version...

Page 21: ...Introduction Overview of the Micro Tech Thermo Fisher Scientific Micro Tech 9101 9201 Reference Manual Rev J 1 3 Figure 1 2 Panel Mounted Version of the Micro Tech...

Page 22: ...il you have read and understood this entire manual The precautions and procedures presented in this manual must be followed carefully in order to prevent equipment damage and protect the operator from...

Page 23: ...ages in excess of 115V or 230V in these areas Avoid high voltage and static electricity around the printed circuit boards WARNING Maintenance procedures should be performed only by qualified service p...

Page 24: ...ent Installation must comply with specifications regarding wire types conductor sizes branch circuit protection and disconnect devices Failure to do so may result in personal injury and or equipment d...

Page 25: ...quirements This section tells you how to complete the hardware installation for your Micro Tech Please go to the appropriate section depending on which model of Micro Tech you purchased field mounted...

Page 26: ...bottom of the enclosure Do not run conduit through the top or sides of the enclosure The Micro Tech should be mounted in a controlled environment and protected shielded from direct sunlight and shoul...

Page 27: ...losure door 3 Route incoming power wiring through a conduit hole at the bottom right of the enclosure For North America locations a certified Nema 4 4X bushing must be used for openings For other loca...

Page 28: ...derside of Enclosure This model of the Micro Tech is designed to be mounted in an instrument panel The instrument panel should not be exposed to excessive vibration heat or moisture The front bezel wh...

Page 29: ...wires 2 Wire the safety ground to the terminal labeled E on the Power Input Terminal 3 Wire the 24VDC to the terminal labeled on the Power Input Terminal 4 Wire the 24VDC Common to the terminal label...

Page 30: ...l output representing total amount of material that has passed the weighbridge Audit trail Automatic zero and span calibration Auto zero tracking Several software options that may be turned on by keyp...

Page 31: ...cessor Optional functions are automatically turned on when the relevant hardware is installed or after the operator has selected them through the keypad Setup of the Micro Tech is easy and is performe...

Page 32: ...an be delayed to avoid intervention in case of short time peaks Each individual alarm can be programmed to operate as alarm shut down or ignored Two LEDs indicate the cumulative status of alarms and s...

Page 33: ...Tech in this case The Remote I O is a typical master slave communication where the main PLC is the master or scanner and the remote devices are slaves or adapters Ethernet Port The Micro Tech has a b...

Page 34: ...t length When the required number of pulses is reached the instrument ends the calibration test and compares the actual totalized value to the theoretical one 0 for zero calibration and calculates the...

Page 35: ...ro calibration as long as the feed rate remains below AZT preset value The Zero Tracking function is limited to a maximum value of Deviation that is set as a percent of full scale in the Setup scroll...

Page 36: ...nformity with the Low Voltage LVD Directive and Electromagnetic Compatibility EMC Directive has been demonstrated using harmonized European Norm EN standards published in the Official Journal of the E...

Page 37: ...ent Control and laboratory use Part 1 General requirements The Micro Tech 9000 series has been tested with the EN and IEC standards listed below IEC EN 61326 1 Electrical equipment for measurement con...

Page 38: ...ient Storage 22 to 158 F 30 to 70 C Operating 4 to 140 F 20 to 60 C Relative Humidity Maximum relative humidity 80 for temperatures up to 31 C decreasing linearly to 50 humidity at 40 C Pollution Degr...

Page 39: ...acts recommended Frequency range Voltage current type sensor 0 25 to 2 0 kHz Contact closure type sensor 0 25 to 30 Hz Low threshold 1 3 VDC min High threshold 2 2 VDC max Low or High Pulse Duration V...

Page 40: ...o Fisher Scientific equipment only RS 485 2 and 4 wire multi drop in RS 485 Data rate 110 to 19200 bits second operator selectable from the keypad Data Format Asynchronous bit serial selectable parity...

Page 41: ...and span calibrations of the scale Your Micro Tech is then ready to go into operation There are five basic steps in the initialization process as follows Determine the correct belt scale code of the c...

Page 42: ...t scale code of 49 for example there are only 11 required parameters In contrast if you have a pivoting scale with a belt scale code of 1 for example there are 18 required parameters So knowing your b...

Page 43: ...r and scale you are using at your particular facility You will need this information when you initialize the Micro Tech software and can save time and effort by gathering this conveyor data now rather...

Page 44: ...when you come to initialize the Micro Tech software Table 2 1 Initialization Data Sheet Non Pivoting Parameter Details of Your Particular System Defaults from Table A 1 or other Number of weigh idler...

Page 45: ...lization Data Sheet Pivoting Parameter Details of Your Particular System Defaults from Table A 1 or other Pivot to load cell distance inches 32 Number of weight idlers 1 Pivot to first idler distance...

Page 46: ...parameters listed above please see pages 2 5 through 2 10 Measure the distance from the pivot to the load cell and enter the result in Table 2 1 or Table 2 2 Figure 2 1 Pivot to Load Cell Distance Cou...

Page 47: ...tion is installed measure the height from the pivot to the test weight and enter the result in Table 2 1 or Table 2 2 If the test weight is below the pivot the value is negative If this option is not...

Page 48: ...2 2 If this option is not available leave at the default value Figure 2 6 Pivot to Carriage Height If the static weight option is installed measure the height from the carry roller to the conveyor str...

Page 49: ...the result in Table 2 1 or Table 2 2 An easy way to arrive at a measurement in degrees for angle is to measure the length AC divide by length AB and look up the cosine Figure 2 9 Conveyor Angle If th...

Page 50: ...nveyor system is equipped with one or in some special instances two optional speed sensors An optional speed sensor feeds very precise conveyor speed readings to the Micro Tech If you did not order an...

Page 51: ...inutes To help you work through the necessary calculations here is a real life example We have just finished installing a 56ft long conveyor at our facility that has a belt length of 120 feet that is...

Page 52: ...ds 5 Now it s your turn to collect your own data and enter it into Table 2 1 or Table 2 2 You are now going to collect data from your own particular conveyor system to determine the Test Duration Here...

Page 53: ...o Tech all the data you entered in Table 2 1 or Table 2 2 Please see page 3 1 of the manual to familiarize yourself with the Micro Tech s console and how to use it This section gives you step by step...

Page 54: ...ech s RAM memory is erased and everything is returned to its initial start up state As a result cold starts are used for example to restore all the Micro Tech settings from a previously made back up f...

Page 55: ...numbers for day If you make a mistake press the Clear button We entered 21 for day as shown below NOTE The Micro Tech displays the date in the month day year format and requires two numbers in the mon...

Page 56: ...ttons on start up scrolling through the fields again to get to this screen Either way make sure this screen is displayed before proceeding 7 You are now ready to enter the correct time as described be...

Page 57: ...ress the Enter button 5 Press the down arrow button to move to the minute field 6 Press the Edit button The Micro Tech clears the minute entry field leaving just the underline 7 Use the keypad to ente...

Page 58: ...Micro Tech to restore your previously saved settings this is where you would insert the back up flash drive into the USB port 11 After waiting briefly press the down arrow button to bring up the langu...

Page 59: ...u are using with your Micro Tech The default value for the Micro Tech is determined by the belt scale code 1 The Micro Tech display should currently look like this 2 The Micro Tech menus are also know...

Page 60: ...ose the Two A D channels option Review the system specific wiring diagram to determine the number of load cells on your scale The Type of scale scroll is not available on a model 9201 Micro Tech 5 Pre...

Page 61: ...Mixed option 1 The Micro Tech display should currently look like this 2 The default selection for Measure Units depends on which Language was selected initially 3 To choose a different selection Engl...

Page 62: ...ou made in previous topic Selecting English Metric Units The Micro Tech display should currently look like this if you chose English units 1 Tons the standard U S ton 2 000 lbs also known as the Briti...

Page 63: ...wn arrow button to bring up the length units screen see the next section This menu allows you to choose the length units used by the Micro Tech Clearly which length units are available in this menu de...

Page 64: ...ess the down arrow button to bring up the rate units screen go to page 2 23 The Micro Tech display should currently look like this if you chose metric units 1 Meters is the default value 2 As no other...

Page 65: ...look like this if you chose English units 1 Standard U S tons equivalent to British short tons per hour Tph is the default value 2 Repeatedly press the Choice button to select other rate units shown...

Page 66: ...1 Metric tonnes per hour t h is the default value 2 Repeatedly press the Choice button to select other rate units shown below then press the Enter button kg mn Kilograms per minute t mn Metric tonnes...

Page 67: ...button to select other rate units shown below then press the Enter button Lb h Pounds per hour Tph Standard tons per hour LTph Long tons per hour kg mn Kilograms per minute t mn Metric tonnes per min...

Page 68: ...this menu are to use either pounds or kilograms The Micro Tech display should look something like this 1 Accept the default value Pounds or kg 2 To choose a different load cell unit for example becaus...

Page 69: ...he menus above 1 To enter the maximum capacity of your particular scale press the Edit button and use the keypad to enter the appropriate value using the decimal point if needed In addition please not...

Page 70: ...results reported to one decimal place that is to the nearest 200 lbs you would choose a scale division of 0 1 As a result hourly rates would be reported as for example 1742 8 Tph tons per hour Please...

Page 71: ...belt scale code for the particular scale you are using in your facility The current weigh bridge configurations offered by Thermo Fisher Scientific as well as the necessary belt scale codes are listed...

Page 72: ...allows you to scroll through a list of parameters to make quite sure they are correct NOTE You must check the conveyor values suggested by the Micro Tech If there is a mismatch between the suggested v...

Page 73: ...t to first idler distance is explained in Figure 2 3 Check the value Press the down arrow button to move on The pivot to test weight height is explained in Figure 2 4 Check the value Press the down ar...

Page 74: ...the button to display a negative sign in front of the number 4 Press the Enter button 5 Press the down arrow button to move on Every load cell has a cord to which is attached a label that displays the...

Page 75: ...like this Nr Number The zero test can be either long or short You should always use the long test when initializing the Micro Tech The long and short tests are assigned the following code numbers Long...

Page 76: ...button at any time returns you to the Establish Test Duration screen 4 Start the belt and when it is running at maximum speed press the Continue button The following screen appears The default value...

Page 77: ...2 for the Number of revolutions for test We entered 5 revolutions as described in our example on page 2 11 8 Press the Enter button and the following screen appears The default value is 30 seconds 9...

Page 78: ...11 10 Your screen should now look something like this 11 Press the Enter button and the following screen appears The time display will start counting down to zero The Micro Tech is now performing the...

Page 79: ...he time in seconds it took to complete the test The data in your screen will of course be different In our example belt length 120 ft So 600 feet 120 x 5 were tested during five revolutions of the bel...

Page 80: ...calibrated message check all the numbers you entered in your Initialization Data Summary table Table 2 1 or Table 2 2 Then go back to the Entering the Appropriate Conveyor Data section above see page...

Page 81: ...his The next step is to go to the Calibrating the Micro Tech section on page 3 5 to perform the initial zero and span calibrations for your scale This is a very important step because the scale will n...

Page 82: ......

Page 83: ...ter describes the Micro Tech console and tells you how to operate and calibrate the Micro Tech on a day to day basis There are four major parts to the Micro Tech console as follows Display screen Keyp...

Page 84: ...a number it operates as an 8 key Similarly in the print menu when you are naming your output repeatedly pressing the 8 key brings up in succession the letters V and W Arrow Keys The up arrow and down...

Page 85: ...ee below is currently being displayed 3 Start the conveyor running and begin loading it up The screen below is known as the Run screen it shows the total tons Tons that have crossed the scale since th...

Page 86: ...ls as follows Master Total This works like the odometer on a vehicle and records the total amount of weight that has crossed the scale since the Micro Tech was first put into operation And like a vehi...

Page 87: ...owing 1 Make sure the Run screen see Figure 3 2 is currently being displayed 2 Press the Totals button and the reset screen appears 3 Press the Reset button and the confirmation screen appears Resetti...

Page 88: ...e calibrated on a daily weekly monthly or other regularly scheduled basis You should run the zero calibration routine often to ensure that the accuracy of the scale is optimized To run a zero calibrat...

Page 89: ...e number of seconds shown in Time remaining is calculated based on the current speed sensor pulse frequency and estimates the time remaining for a complete test 5 The calibration time in seconds that...

Page 90: ...However as there has been a small amount of drift we decide to reset the zero to the newly established zero point 6 Press the Yes button to accept the new zero and the zero changed screen appears 7 P...

Page 91: ...in Menu 1 screen appears 3 Press the Span Calibration button and the following screen appears If the R Cal auto span option shown above is not displayed go to page B 5 to change the span calibration m...

Page 92: ...ime remaining is calculated based on the current speed sensor pulse frequency and estimates the time remaining for a complete test 5 The calibration time in seconds that you established during the Mic...

Page 93: ...ady to put your Micro Tech into operation Performing Any Subsequent R Cal Once you have initialized your Micro Tech and are doing an R Cal as part of your weekly daily or other routing testing proceed...

Page 94: ...on that is do not change the span and the following screen appears Your numbers will of course be different 7 Press the Run soft key in the display to return the Micro Tech to the Run mode 8 Run sever...

Page 95: ...and should require very little maintenance The front panel can be wiped clean with a damp cloth and if necessary a mild detergent never use abrasive cleaners especially on the display window As a pre...

Page 96: ...ting your scale s zero and span How often should I check the mechanical installation of the scale The scale should only need daily preventive maintenance checks for material build up in critical areas...

Page 97: ...so on An adjustable damping filter is provided separately for the displayed rate and current outputs What kind of information is available from the digital output The programmable outputs are as follo...

Page 98: ...enu 3 Diagnostics Two menus pertain to load cell troubleshooting as follows A D Gross and A D Net Diagnostic Scroll 1 Weight on Load Cell Diagnostic Scroll 2 This is a digital representation of the mV...

Page 99: ...particular system Four wire load cell with no field sense leads Four wire load cell with field sense leads Six wire load cell with field sense leads Please refer to the figure below for more informati...

Page 100: ...Jumpers shown below See Appendix A Premium A D Jumper Locations for Model 9201 jumper locations Figure 4 2 Four Wire Load Cell with Sense Leads Please refer to the instructions and figure below when...

Page 101: ...J16 and or J21 5VDC If NO Verify proper position of sense jumpers see page A 28 If YES Proceed to verify load cell mV output see step 3 3 Verify load cell mV output s Is load cell mV output within ra...

Page 102: ...J Thermo Fisher Scientific Figure 4 3 Six Wire Load Cell with Sense Leads This section helps you troubleshoot the Micro Tech s analog input output circuitry There are two types of optional analog outp...

Page 103: ...nt output s Analog voltage inputs An analog signal failure could be caused by the following Failed analog board Bad or incorrect wiring Remote device failure CPU board failure The troubleshooting proc...

Page 104: ...or troubleshooting the Micro Tech s analog input output boards 1 Confirm the DC mA voltage source If GOOD Proceed to Confirm analog output see step 2 A dual analog output board has individual power so...

Page 105: ...ro Tech 9101 9201 Reference Manual Rev J 4 11 Figure 4 5 Analog Troubleshooting Confirm Output 3 Confirm analog loop If GOOD Analog board is performing as expected If BAD A problem exist in either wit...

Page 106: ...for Input 2 5 Verify voltage input Measure input signal 1 5vdc representing 0 100 If GOOD Replace analog input output board If BAD Verify input jumpers J1 and or J2 If correct proceed to Verify outpu...

Page 107: ...erence Manual Rev J 4 13 If BAD Verify field wiring Remote device failure Figure 4 8 Analog Troubleshooting Verify Source Output 7 Verify mA loop circuit Place mA meter in series with analog loop If G...

Page 108: ...gital outputs are controlled by the Micro Tech The LED indicator on the Opto22 module indicates what the integrator is forcing that is Off the LED is off or On the LED is on Here is the procedure to f...

Page 109: ...e activated and the equipment may start 4 Press the Continue button Here is the procedure to follow to test the Opto22 output module 1 Verify the source voltage AC or DC Measure the source voltage If...

Page 110: ...Proceed to View Opto LED see step 4 If BAD Replace fuse 4 View Opto22 LED By forcing the output open and closed does the LED indicate the appropriate logic condition If GOOD Replace the Opto22 output...

Page 111: ...atus that is 0 open 1 closed The display reads right to left that is from Inputs 1 through to Input 5 If your Micro Tech is equipped with an optional digital input output board press the Next button t...

Page 112: ...Reference Manual Rev J Thermo Fisher Scientific 3 View Opto LED With input voltage supplied LED should be lit If GOOD Replace either the Opto22 input module or the Micro Tech CPU board If BAD Replace...

Page 113: ...al eight in out digital DIO board 1 Follow steps 1 2 in the Navigating to the Test Output Screen section on page 4 12 to display the Digital Input Test screen shown below 2 Monitor the open collector...

Page 114: ...Problems 4 20 Micro Tech 9101 9201 Reference Manual Rev J Thermo Fisher Scientific 3 Outputs Sourcing Force the output number open closed Figure 4 13 DIO Troubleshooting Sourcing 4 Outputs Sinking Fi...

Page 115: ...Return button to back step to the previous screen Asterisks indicate an out of range span calculation 1 Verify a load was applied to scale Verify your set up data 2 Press the No button 3 If you press...

Page 116: ...nd request a reconfigure This message is only available for a limited time If the Micro Tech is not reconfigured the option installed will not function The No Speed Pulses warning can occur during a n...

Page 117: ...ng a proper input signal If you are running the Micro Tech in the Simulated mode verify the conveyor s run contact is providing input to the Micro Tech 2 Press the Continue button During certain set u...

Page 118: ......

Page 119: ...detailed list of the spare parts available for your Micro Tech please see the Parts List in Table 5 1 on page 5 3 Before returning the Micro Tech or any other equipment to Thermo Fisher Scientific yo...

Page 120: ...needed parts make sure you have the following information ready Your Micro Tech serial number and belt scale code Your company s purchase order PO number Please note that a hard copy of your PO is req...

Page 121: ...ermofisher com parts bulk us thermofisher com Germany Chile 49 0 208 824930 56 2 2378 5080 49 0 208 852310 fax 56 2 2370 1082 fax service oberhausen thermofisher com Italy China 39 02 959514 1 86 10 8...

Page 122: ...MT9000 102949 PCBA 4 20MA OUTPUT BD 1 CH MT9000 100744 PCBA DC INPUT BD MT9K 100785 PCBA PLANT A D BD MT9000 102450 PCBA OPTO 22 INPUT MT9000 102999 PCBA OPTO 22 OUTPUT BD MT9000 103003 PCBA RELAY OUT...

Page 123: ...her Scientific Micro Tech 9101 9201 Reference Manual Rev J A 1 Appendix A Additional Installation Information The following page shows a copy of a typical System Data Sheet for the Micro Tech System D...

Page 124: ...A 2 Micro Tech 9101 9201 Reference Manual Rev J Thermo Fisher Scientific...

Page 125: ...Thermo Fisher Scientific Micro Tech 9101 9201 Reference Manual Rev J A 3 Here is a copy of a typical Door Label for the Micro Tech Door Label...

Page 126: ...0 0 42 5 36 4 5 6 5 7 3 11 10 22 18 36 50 30 1 2 62 54 18 0 0 0 0 38 36 4 75 6 5 6 5 3 12 10 22 18 36 50 30 1 2 71 63 21 0 0 0 0 42 42 4 75 6 5 6 5 3 13 10 22 18 36 50 30 1 2 80 72 24 0 0 0 0 48 48 4...

Page 127: ...4 4 0 0 0 0 0 0 0 0 39 37 0 0 0 3 52 10 14 4 50 14 4 4 0 0 0 0 0 0 0 0 47 24 0 0 0 3 53 10 30 50 30 1 1 0 0 0 0 0 0 0 0 36 0 0 0 1 8 54 10 11 18 42 WTS 1 1 55 50 48 0 0 0 0 0 40 36 6 5 6 5 6 5 3 55 10...

Page 128: ...4 86 61 23 62 70 87 23 62 70 87 0 0 86 61 47 24 4 5 6 5 7 3 201 1 1 31 89 24 02 0 0 0 0 0 24 02 39 37 16 81 6 38 4 92 3 202 1 1 31 89 24 02 0 0 0 0 0 24 02 39 37 16 81 4 41 4 92 3 203 1 1 31 89 24 02...

Page 129: ...7 4 72 3 403 1 1 39 37 29 53 29 53 0 0 0 0 29 53 39 37 4 06 4 17 4 72 3 404 405 2 1 0 0 0 0 0 0 0 0 39 37 0 0 0 2 406 4 2 0 0 0 0 0 0 0 0 39 37 0 0 0 3 407 4 3 0 0 0 0 0 0 0 0 47 24 0 0 0 3 408 4 4 0...

Page 130: ...ivolts volt Figure 2 10 This procedure calibrates the belt to the system and establishes a test duration for all the simulated calibrations A test duration can be acquired or entered manually Acquirin...

Page 131: ...croll 12 screen shown below To use the partial belt length method do the following 1 Press the Acquire button and the following screen appears 2 Mark and measure a section of the belt The section you...

Page 132: ...al length of the section you marked The default value is 200 feet 5 Press the Enter button We entered 300 feet 6 Press the Start button BATCH Ent len between two marks on belt Length 200 0 ft READY AL...

Page 133: ...t The belt has made at least three complete revolutions The total running time of the test exceeds six minutes 9 After the last revolution has been counted press the Done button BATCH Press COUNT each...

Page 134: ...eturns you to the Cal Data Scroll 12 screen shown below To use the full belt length method do the following 1 Make a chalk mark on the belt 2 Press the Acquire button and the following screen appears...

Page 135: ...enter the entire length of the belt Then press the Enter button 5 Press the Start button BATCH Ent len of one belt revolution Length ft READY ALARM CALIB ABORT ENTER CLEAR BATCH Ent len of one belt re...

Page 136: ...are met The belt has made at least three complete revolutions The total running time of the test exceeds six minutes 8 After the last revolution is counted press the Done button BATCH Press COUNT each...

Page 137: ...ptional item Test chain Optional item Other Material Factoring pre or post weighed material User supplied Since all Micro Tech models are equipped with R Cal this example will assume an R Cal span It...

Page 138: ...arrow button repeatedly until the Cal Data Scroll 10 screen appears 4 Press the Next button and the Weights screen appears BATCH START OF SCROLL Use SCROLL keys to view selections READY ALARM CALIB BA...

Page 139: ...k like this The auto span procedure determines the percent static weight span error 1 Press the Menu button to bring up the Main Menu 1 screen Static Weight BATCH CAL DATA SCROLL 10 Material FACTOR We...

Page 140: ...n appears The auto span procedure determines the percent of scale weight span error 1 Add your test weight to the conveyor Auto Span BATCH AUTO SPAN Weights Press START to begin Weight calibration REA...

Page 141: ...creen appears 4 When the Auto Span procedure is completed the span percent error is displayed Record the results and note whether the error is positive or negative Record Results BATCH AUTO SPANNING T...

Page 142: ...pears 2 Press the Run button and the following screen appears 3 Press the Run button and the Micro Tech Run screen reappears as shown below BATCH SPAN UNCHANGED New span xxxxxx Old span xxxxxx READY A...

Page 143: ...arrow button and the Material Factor screen appears 6 Press the Next button and the Weights screen appears BATCH MAIN MENU 2 Press MENU for more READY ALARM CALIB SCALE DATA DISPLAY CALIB DATA BATCH...

Page 144: ...enter the weight error Then press the Edit button The error you entered is displayed in the screen below 8 Press the Run button and the Micro Tech Run screen reappears as shown below BATCH CAL DATA SC...

Page 145: ...7 11 12 13 14 15 16 17 21 22 23 24 25 26 31 32 33 34 35 36 37 38 41 42 43 44 45 46 47 48 LOADCELL 1 LOADCELL 2 COMM B COMM A SPU and PULSE OUT J16 Load Cell 1 1 SHIELD EARTH 2 EXCITATION 3 EXCITATION...

Page 146: ...5 A RX 24 RS 485 B RX 25 COMMON 26 SHIELD EARTH J37 COMM A Isolated RS 485 232 31 RTS 485 OUT 32 TXD 485 OUT 33 RXD 34 485 IN 35 485 IN 36 CTS DCO 37 UART GND ISOLATED 38 SHIELD EARTH J29 Speed signal...

Page 147: ...ock definitions for the Premium A D Board are shown below Model 9201 Micro Tech only J9 Premium A D Load Cell 1 Shield Earth GND 2 Signal COM 3 Excitation 5v 4 Excitation 5v 5 Excitation Sense 6 Excit...

Page 148: ...A 26 Micro Tech 9101 9201 Reference Manual Rev J Thermo Fisher Scientific Here are the jumper locations the Micro Tech motherboard Motherboard Jumper Locations...

Page 149: ...Thermo Fisher Scientific Micro Tech 9101 9201 Reference Manual Rev J A 27 Here are the Load Cell jumper locations of the Premium A D board Model 9201 Micro Tech only Premium A D Jumper Locations...

Page 150: ...r the user or by field service personnel J2 Jumper Setting Description Installed Core voltage applied to pin 25 of BDM JTAG header J2 Not Installed This jumper position must not be altered by either t...

Page 151: ...se Selection Jumpers Jumper Jumper Settings Default J14 J15 Jumper Setting Description Pins 1 2 Local Sense Channel 1 Pins 2 3 Remote Sense Channel 1 Not Installed Remote Sense Channel 1 J14 1 2 Insta...

Page 152: ...stalled Remote Power J24 Installed J25 Installed J26 J27 Speed input 2 Output 5 Jumper Setting Description Installed Local 24V Power Not Installed Remote Power J26 Installed J27 Installed J28 J30 J28...

Page 153: ...J39 Installed Pins 1 2 J38 COMM A UART 0 Jumper Setting Description Pins 1 2 RS 485 Receive U0RXD Pins 2 3 RS 485 Receive U0CTS J38 Installed 1 2 J40 COMM A Termination UART 0 Jumper Setting Descripti...

Page 154: ...stalled 120 termination J46 Not Installed Here are the jumper settings for the Premium A D board Model 9201 Micro Tech only Premium A D Board Load Cell Excitation Sense Jumper Jumper Settings Default...

Page 155: ...ifications Model 9101 Load Cell Excitation Power Supply 5 VDC 10 90 mA minimum load impedance 58 ohms Output short circuit 0 5 A maximum Load Cell Load cell input circuits 2 each Number Up to six 6 35...

Page 156: ...3000ft 914m with sense Sensitivity 0 5mV V to 3 5 mV V keypad selectable Load cell input circuit Input Impedance 1M ohm minimum Maximum Usable Signal 114 of 3mV V Internal A D counts 3mV V 7 341 000...

Page 157: ...e dry contact inputs on the optional DC Input Board or three 3 programmable opto 22 inputs modules on the optional Opto22 Input Board Eight 8 programmable inputs on the optional Digital I O 8in 8out B...

Page 158: ...rd Install in expansion slot J6 on the motherboard Use UL 1015 wire 16AWG 1 sq mm or smaller This is an optional board with three inputs inputs 3 5 Type Current sourcing to common ground Designed for...

Page 159: ...VAC In Module G4IAC5A Input voltage range 180 280 VAC or VDC Input current at maximum line 5mA 120VAC In Module G4IAC5 Input voltage range 90 140 VAC or VDC Input current at maximum line 5mA 32VDC In...

Page 160: ...n the motherboard Use UL 1015 wire 16AWG 1 sq mm or smaller This is an optional board Four 4 outputs outputs 1 4 Panel version Rated 33 VAC at 2A Fusing requirement 3A Rated 70 VDC at 0 5A Fusing requ...

Page 161: ...tput board The module options are as follows 240VAC Out Module G4OAC5A Output voltage range 24 280 VAC 2A at 50 C ambient Replaceable 250V 4A fuse 60VDC Out Module G4ODC5 Output voltage range 5 60 VDC...

Page 162: ...volt current sinking or current sourcing isolated outputs Output current must be limited to 80 mA maximum continuous The inputs and outputs are powered by an external 24 VDC power source The isolated...

Page 163: ...onal board Install in one of the motherboard expansion slots J10 J13 Single channel high level current output Outputs Rate Speed Load Optically isolated Isolated power source Voltage output by adding...

Page 164: ...ev J Thermo Fisher Scientific Board Diagram Part number 100744 This is an optional board Install in one of the motherboard expansion slots J10 J13 Inputs Incline compensation Moisture compensation Out...

Page 165: ...play Volts 0 to 5 V 1 to 5 V Jumpers J1 and or J2 are used to select 250 ohm resistance for the Current inputs Input impedance 100 k nominal differential Maximum usable input voltage 106 of full scale...

Page 166: ...model 9101 only Each load cell channel provides its own buffer amplifiers for driving the A D converter IC s differential reference voltage from the excitation sense voltage resistive divider The loa...

Page 167: ...installed jumper do not move Load Cell Sense Jumper Defaults Jumper Four Wire Load Cell internal sense Six Wire Load Cell external sense J14 1 2 2 3 J15 1 2 2 3 J19 1 2 2 3 J20 1 2 2 3 Factory option...

Page 168: ...and 4 wire multi drop in RS 485 20 mA loop is passive ONLY Interfacing RS 485 supports 2 wire or 4 wire multi drop networking RS 232C provides support for modem Data rate 300 to 19200 operator selecta...

Page 169: ...are shown below Jumpers Mode OP1 OP2 OP3 OP4 OP5 OP6 RS 232 A A A A A B RS 485 B A B B MDP TRM 20 mA B B A A A C Default MDP TRM For RS 485 only For RS 485 only OP5 OP6 A Normal A Terminated B Multi...

Page 170: ...us is a typical master slave communication where the main PLC is the master or scanner and the Micro Tech device is a slave The connection is EIA RS 485 through a 2 wire twinax Siemens cable Profibus...

Page 171: ...rface board contains the Siemens SPC3 Profibus DP controller ASIC The SPC3 controller is an integrated circuit provided by Siemens that handles the interface between the Micro Tech slave and the maste...

Page 172: ......

Page 173: ...reens Each time you press the Menu button the next menu screen appears 1 To start your overview press the Menu button to bring up the Main Menu 1 screen This screen contains the following functions ZE...

Page 174: ...TION Data Allows you to choose which calibration method to use and set up material factors and auto zero tracking AZT 3 Press the Menu button to bring up the Main Menu 3 screen This screen contains th...

Page 175: ...he Micro Tech s digital and analog inputs outputs ALARMS Allows you to define how alarms are generated ETHERNET Allows you to set the Micro Tech s EtherNet communication parameters 5 Press the Menu bu...

Page 176: ...ns AUDIT Trail Allow you to capture changes to the Micro Tech settings and to record the details of all Micro Tech calibrations LINEARIZATION Applies a factor to the span settings at multiple flow rat...

Page 177: ...load consists of the scale carriage idler s and the section of belt within the weigh area The zero number displays the load cell signal mVDC in digital form Here is the Zero Calibration menu table Par...

Page 178: ...stment to the Zero Number AZT Shows the total deviation of the AZT correction factor from the Zero Reference value AZT AZT The Micro Tech has three calibration methods as follows The electronic R Cal...

Page 179: ...to bring up the Calibration Data screen 5 Press the Choice button repeatedly until the method you want to use appears in the screen as shown below BATCH MAIN MENU 2 Press MENU for more READY ALARM CAL...

Page 180: ...ch calibration R Cal weights or chain method you chose above Please refer to the appropriate screen below R Cal Method Test Chain Method BATCH CALIB DATA SCROLL 1 Calibration mode Weights READY ALARM...

Page 181: ...the link at the end of the chain Press the Edit button and use the keypad to enter the appropriate value Static Weights Method Add up the total weight you added to the scale press the Edit button and...

Page 182: ...escribed above 2 Make sure the conveyor belt is stopped before applying the weights or chains CAUTION Do not apply weights or chains until instructed to do so by the Micro Tech 3 When using the weight...

Page 183: ...d load method has no material factor installed Ignore this screen and press Exit Automatic Material Factoring should not be used If you want to use the manual Material Factoring procedure see page A 1...

Page 184: ...complete be sure to remove the Chain or Weights before returning to normal operation BATCH AUTO SPAN COMPLETE Error X XX Change span READY ALARM CALIB NO YES ADV BATCH REMOVE weight before returning t...

Page 185: ...RUN REPEAT ADV XXXX Old span Error Total Calcon Mat l fact Only present when a Material Factor is installed for the current simulated load method XXXX New span SPAN UNCHANGED RUN REPEAT ADV XXXX Old...

Page 186: ...on RUN MENU AUTOSPAN COMPLETE EXIT FACTOR REPEAT Note Only available if material calibration is done before and the current simulated load method has no material factor installed Error Unfactored Calc...

Page 187: ...weighing NOTE Perform zero and span calibrations prior to Material Testing Material testing requires sufficient material to run without interruption continually for the required tonnage and time of 6...

Page 188: ...s the down arrow button and the Start screen appears 4 Make sure that the belt is running empty then press the Start button BATCH START OF SCROLL Use SCROLL keys to view selections READY ALARM CALIB B...

Page 189: ...Run screen appears Your figures will of course be different 7 When you have completed the test press the Done button and the following screen appears 8 Press the No button meaning your reference weigh...

Page 190: ...ight screen shown below 10 Press the Edit button and use the keypad to enter the exact reference weight 11 Press the Enter button and the following screen appears BATCH xxx xx Tons Enter reference wei...

Page 191: ...the Factor button the Micro Tech uses a simulated calibration to populate the parameters in the Material Factor function 14 Press the No button and the Run screen reappears Here is the Material Calibr...

Page 192: ...Change span SPAN CHANGED RUN MENU FACTOR This option is only available if a simulated Auto Span calibration has been completed beforehand New span Old span MAT L CALIBRATION RCAL WTS CHAIN Each optio...

Page 193: ...vice 2 Totalization Units Tons LTons Pounds Kg tonnes Password Service 3 Length Units Feet meters Selects unit of measurement Belt length Idler spacing Scale lever ratios static weight measurements Pa...

Page 194: ...ISPLAY SCROLL 10 Run Display Line 3 setting 14 L C Fail Time Delay 2 2 90 sec This timer starts when a load cell signal fail alarm has been generated and after the timer has elapsed the software force...

Page 195: ...1 6 Number of Idlers mounted on the scale carriage Default value for Belt Scale code 1 shown 4C Pivot to 1st idler distance 24 00 0 5000 in On a pivoting scale measurement from pivot to 1st idler cen...

Page 196: ...to carriage height centerline Note Scale code dependent Default value for Belt Scale code 1 shown 4O of load cells 2 1 6 Quantity of load cells wired to the Micro Tech Default value for Belt Scale cod...

Page 197: ...eed input Single dual simulated Informs the Micro Tech where the speed input is physically wired or not wired in the case of simulated 11 Zero dead band range 0 0 0 5 Low level threshold where Rate an...

Page 198: ...is scroll displays the date of the last calibration and the expected date of the next one based on the entry in the previous screen 10 Material FACTOR R CAL 0 00 99 99 99 99 Correction factor for Rcal...

Page 199: ...bled under the PRINT scroll Generates an alarm when the change from the Zero Reference value to the value of zero with new factor applied exceeds the max deviation 17 Auto zero tracking Warm Time 1 0...

Page 200: ...t loading remains under 0 12 of full scale belt loading for a period of time equivalent to one belt revolution The indicator is the green ZERO LED above the display The function can be activated only...

Page 201: ...you have already created a password in the Diagnostic Menu There are two passwords Service and Operator There are three protection levels with specific related passwords Press the Protection button a...

Page 202: ...desired level as described below Decreasing the protection level requires a password 1 Begin by selecting a protection level lower than the current Press the Limited soft key 2 The display will promp...

Page 203: ...talled the following screen is displayed You can enter either the Operator or the Service password However if you enter the Operator password and the variable or function requires the Service password...

Page 204: ...re is the Protection menu table Parameter Selection Description PROTECTION LEVEL NONE LIMITED PROTECTED The PROT menu is activated when a Service or Operator password is entered Sets protection level...

Page 205: ...v zero set Factory adjusted Not to be changed in the field 2B Loadcell output span 0 30000 Fine adjust of load cell mv v span set Factory adjusted Not to be changed in the field 3 Prescale___ 1_____ p...

Page 206: ...DP Communication A AD BOARD 2 channels Defines option pcba in bus slot 1 The system automatically recognizes when optional boards are installed When a board is acknowledged the related information st...

Page 207: ...number of DIO 8in 8out boards installed Shows the status of each digital output and allows the operator to force the output for testing purposes The output when forced stays on until the CLEAR soft ke...

Page 208: ...e Profibus board is installed 10 11 12 13 Keyboard test Press keys Hit RUN key twice quickly to exit The USB menu allows you to download all the Micro Tech settings to a flash drive to back them up an...

Page 209: ...ollowing commands format x fs fat16 a 512 Press the Enter key where x is the actual drive letter assigned to the USB device 4 Answer Yes to any warnings about erasing all data on the device 5 When com...

Page 210: ...t to use this menu 1 Press the USB button and the following screen appears 2 Press the down arrow button and the USB configuration screen appears Uploading and Downloading Micro Tech Settings BATCH MA...

Page 211: ...h Drive To save the Micro Tech s settings to a flash drive do the following 1 Press the Save button and the following screens displays the progress of the save operation BATCH USB Configuration Archiv...

Page 212: ...NOTE Be careful using the Restore procedure because all your current settings will be overwritten 2 Press the Restore button and the following screens displays the progress of the restore operation B...

Page 213: ...xt editor like Notepad or other similar program The Micro Tech saves the print reports with an 8 dot 3 filename Example 01ADAZP1 SET The filename is defined as follows USB File Name Location In File N...

Page 214: ...to Zero n is scale number AZT Auto Zero Tracking ATn Auto Zero Tracking n is scale number Year Month Code Table CODE MONTH YEAR CODE MONTH YEAR CODE MONTH YEAR CODE MONTH YEAR CODE MONTH YEAR AA Jan 2...

Page 215: ...Jun 2017 DN Aug 2019 EN Oct 2021 AO Mar 2013 BO May 2015 CO Jul 2017 DO Sep 2019 EO Nov 2021 AP Apr 2013 BP Jun 2015 CP Aug 2017 DP Oct 2019 EP Dec 2021 AQ May 2013 BQ Jul 2015 CQ Sep 2017 DQ Nov 201...

Page 216: ...Input Output Define menu table No Parameter Selection Description 1 Current output define 1 2 3 4 Rate off load speed This scroll is only available when a I Out card or a 2I 2O card is installed 1A Cu...

Page 217: ...ding to total Use the CHOICE key to select the action of the analog input use ENTER to confirm Press the CALIB softkey to calibrate 2A Conveyor low posit Moisture calib Low 0 00 Degrees 35 35 0 0 M 0...

Page 218: ...osed NC status of the input By assigning a function to 0 the function is disabled The CommA Scroll 5 Protocol port 1 must be set to Printer or a COMM board must be installed for this option to be show...

Page 219: ...M mode 5 Only available if Loadout card installed High rate2 0 NO Low rate2 0 NO High speed3 0 NO Low speed3 0 NO Totalizer 0 NO Print ready4 0 NO Load WTS 0 NO Dev Alarms4 0 NC Out of range4 0 NO Ze...

Page 220: ...isplayed instead of the New button The Next button is used to scroll between pending alarms Here is a list of alarms 1 Clock Fail The system has detected a failure on the clock calendar circuit Go to...

Page 221: ...e of the Scale Capacity Value 1A Low rate set ___ ___ 10 0 105 10 sec 0 90sec 50 0 Tph 0 525 Tph 1B High rate set ___ ___ 100 0 150 10 sec 0 90sec 500 0 Tph 0 750 Tph 2 Load alarm no yes Sets an alarm...

Page 222: ...Memory checksum during the internal periodic test The RAM is used to store variables and set up data Replace the motherboard 8 ROM fail ALARM SHUT DOWN AUTO RES NONE The system has detected an error...

Page 223: ...this alarm occurs The purpose is to remind the operator that the calibration has not been checked for a considerably long period 37 Overflow Totalizer ALARM SHUT DOWN AUTO RES NONE Only present when t...

Page 224: ...FIB comm error ALARM SHUT DOWN AUTO RES NONE This scroll is only available if the optional Profibus board is installed The following two conditions activate the alarm The Siemens SPC3 Controller insta...

Page 225: ...herboard has one serial channel that can be configured using jumpers as an RS 232 or an RS 485 channel The serial channel can be used for printing or for serial communication with an intelligent devic...

Page 226: ...port 2 No Parity even parity odd parity Note Only available when COMM board is installed 10 Stop bits port 2 1 2 Note Only available when COMM board is installed 11 Word length port 2 8 7 Note Only a...

Page 227: ...a protocol other than Printer is selected for either Protocol port 1 or Protocol port 2 Not damped shows the raw value without any damping without the scale division applied Damped applies the damping...

Page 228: ...layed This section tells you how to use the Micro Tech s print function 1 In the Run screen shown above press the Print button For this to work you must have already set the communications parameters...

Page 229: ...perator and Reset totals Trails Prints out Audit Trails Press the Edit button to select the number of trails to print Press the More button to access more records 3 Press the Print button to print the...

Page 230: ...ready to receive data Password Service 2 End of line CR LF CR LF Different printers use different end of line patterns Select the one you need for your printer Password Service 3 Delay end of line 0 0...

Page 231: ...ult 1 default 2 user defined 9A Number of strings 1 1 3 This scroll is only available when user defined is selected for Total report format 9B Contents string 1 String 1 9C Position string 1 X ___ Y _...

Page 232: ...date of the change All changes can be printed when required To view Audit Trails press the Edit button and using the keypad enter the trail number from 1 to 999 that you want to view Then press the En...

Page 233: ...00000 for the factor the rate is not corrected in that portion of the range A factor lower than 1 000 reduces the span while a factor larger than 1 000 increases the span The ACQ key automatically acq...

Page 234: ......

Page 235: ...S232C RS 485 jumper selectable isolated Comm B is a RS 485 only non isolated RS423 RS232C For point to point asynchronous bidirectional communications maximum 50 feet 15 m Modem capability RS485 RS422...

Page 236: ...J40 J42 J43 RS 232 1 2 1 2 1 2 1 2 1 2 3 4 1 2 1 2 RS 485 4 wire 2 3 1 2 2 3 2 3 1 2 3 4 1 2 1 2 2 wire 2 3 2 3 2 3 2 3 1 2 3 4 1 2 1 2 multi drop 2 3 Transmit with hardware hand shake 2 3 2 3 2 3 2...

Page 237: ...n the mother board to one of the standards as listed in the previous paragraph However only the RS485 RS422 standard allows multi drop communications RS232C can only be used in point to point mode Thi...

Page 238: ...processed only if the address matches the one specified in the set up data of the Micro Tech otherwise it is ignored When the system receives a message the integrity of the message is checked An answe...

Page 239: ...o not generate an answer When a COMM line is set up for printing the system prints out data in different ways depending on the set up Print menus become visible which allow the user to define when and...

Page 240: ...efore 100 ms have passed since the previous one it is lost The data is received by an interrupt routine while it is processed in a task that is executed each 100 ms If the Master sends variables to be...

Page 241: ...e written Write is subject to protection control Low limit The minimum acceptable value for the variable Lower values are considered as errors High limit The maximum acceptable value for the variable...

Page 242: ...Specific RW 0 100 limited Integer 1 AZ Deviation Scale Specific R 0 10000000 none Long 2 Zero Number Scale Specific R 0 100 none Long 2 Zero Reference Scale Specific R 0 120000 Service Long 2 Set Res...

Page 243: ...l Double Scale Specific R 0 0 none Double 4 Reset Total Double Scale Specific R 0 0 none Double 4 Operator Total Double Scale Specific R 0 0 limited Double 4 Batch Total Double Scale Specific R 0 0 li...

Page 244: ...ut Set Res Out Span Number Indicate the span in use in the scale Scale Capacity The scale capacity of the scale Entered by the user at first start up should never be altered Speed Capacity The speed c...

Page 245: ...acity value Entered or downloaded by the user Low Speed Set The set point for the alarm of low speed in percent of the Belt Speed Capacity value Entered or downloaded by the user High Load Set The set...

Page 246: ...fourth row Register Bit Description Status LED 15 Not Used 14 Not Used 13 Not Used 12 Not Used 11 Not Used 10 Not Used 9 Not Used 8 Not Used 7 Not Used 6 Not Used 5 Not Used 4 Zero Ready Label ZERO L...

Page 247: ...t Used Status 2 15 Not Used 14 Not Used 13 Not Used 12 Not Used 11 Not Used 10 Not Used 9 Not Used 8 Not Used 7 Not Used 6 Not Used 5 Not Used 4 Not Used 3 See Other Table 2 See Other Table 1 See Othe...

Page 248: ...h Is Running At Low Rate 0 0 1 1 Batch Is going To Start Wait Batch Is In start Delay Time 0 1 0 0 Batch Is In Coast 0 1 0 1 Batch Hold Suspended at High Rate 1 0 1 0 Batch Hold Suspended at Low Rate...

Page 249: ...e 1 Alarm 3 15 Calibration Time Elapsed Scale 2 14 External Alarm 1 13 External Alarm 2 12 External Alarm 3 11 OVERFLOW TOTALIZER 10 Overflow Totalizer Scale 1 9 Overflow Totalizer Scale 2 8 AZT OVER...

Page 250: ...12 A D Shutdown Channel 3 11 A D Shutdown Channel 4 10 Not Used 9 Not Used 8 Not Used 7 Not Used 6 Not Used 5 Not Used 4 Not Used 3 Not Used 2 Not Used 1 Not Used 0 Not Used Alarm 6 15 Not Used 14 No...

Page 251: ...2 10 Input 3 Digital 8In 8Out Board 2 9 Input 2 Digital 8In 8Out Board 2 8 Input 1 Digital 8In 8Out Board 2 7 Not Used 6 Output 5 Motherboard Output 5 Output 4 Digital Input Board 4 Output 3 Digital I...

Page 252: ...Not Used 0 Not Used I O 5 15 Not Used 14 Not Used 13 Not Used 12 Not Used 11 Not Used 10 Not Used 9 Not Used 8 Not Used 7 Not Used 6 Not Used 5 Not Used 4 Not Used 3 Not Used 2 Not Used 1 Not Used 0 N...

Page 253: ...Not Used 1 Not Used 0 Not Used I O 8 15 Not Used 14 Not Used 13 Not Used 12 Not Used 11 Not Used 10 Not Used 9 Not Used 8 Not Used 7 Not Used 6 Not Used 5 Not Used 4 Not Used 3 Not Used 2 Not Used 1 N...

Page 254: ...d 3 Not Used 2 Not Used 1 Not Used 0 Not Used I O 11 15 Not Used 14 Not Used 13 Not Used 12 Not Used 11 Not Used 10 Not Used 9 Not Used 8 Not Used 7 Not Used 6 Not Used 5 Not Used 4 Not Used 3 Not Use...

Page 255: ...n needs a delay of 0 5 seconds In case a R Cal remote calibration is aborted the calibration running flags will turn off after this delay See bit Confirm Zero Span to save the calibration speed signal...

Page 256: ...running function speed signal 8 3 8 Batch Standby 8 Temporarily stops a Batch Load Out sequence Batch can resume if a Start command is sent later Batch Enabled 4 2 4 Batch Abort 4 Stops the Batch Load...

Page 257: ...o Tech as shown in the table below Host PLC Slave Micro Tech STX 02H DLE 10H HEADER HI 00H HEADER LO 00H COMMAND HI 41H COMMAND LO 44H DESTINATION DB XXH BYTE COUNT HI 00H BYTECOUNT LO XXH no CF FFH a...

Page 258: ...cro Tech Host PLC Slave Micro Tech STX DLE 10H HEADER HI 00H HEADER LO 00H COMMAND HI 45H COMMAND LO 44H DESTINATION DB 00H DESTINATION DW XXH BYTE COUNT HI 00H BYTE COUNT LO XXH no CF FFH all CPUs FF...

Page 259: ...Manual Rev J C 25 Host PLC Slave Micro Tech BCC XXH DLE 10H Example 3 Host Wants to Write Batch Setpoint Set batch 100 Tons Host PLC Slave Micro Tech 02H 10H 00H 00H 41H 44H 00H 6CH 00H 02H FFH FFH 4...

Page 260: ...o Fisher Scientific Host PLC Slave Micro Tech 10H 03H 13H 10H Example 4 Host Wants to Receive Batch Setpoint Batch setpoint 100 Tons Host PLC Slave Micro Tech 02H 10H 00H 00H 45H 44H 00H 6CH 00H 02H F...

Page 261: ...odel 9101 9201 Register Address Notes success_flag 20 H 0 successful 1 failed display 1 21 H display 2 22 H display 3 23 H display 4 24 H display 5 25 H display 6 26 H display 7 27 H display 8 28 H di...

Page 262: ...27 3B H display 28 3C H display 29 3D H display 30 3E H display 31 3F H display 32 40 H display 33 41 H display 34 42 H display 35 43 H display 36 44 H display 37 45 H display 38 46 H display 39 47 H...

Page 263: ...total 2 67 H operator_total 1 68 H operator_total 2 69 H batch_total 1 6A H Load Out option required batch_total 2 6B H Load Out option required batch_set_point 1 6C H Load Out option required batch_s...

Page 264: ...set 1 82 H low_load_set 2 83 H reg_set_point_1 1 84 H reg_set_point_1 2 85 H reg_set_point_2 1 86 H reg_set_point_2 2 87 H dsp_set_point_1 1 88 H dsp_set_point_1 2 89 H dsp_set_point_2 1 8A H dsp_set_...

Page 265: ...1 H AZT step 1 A2 H AZT step 2 A3 H Belt warm up time 1 A4 H Belt warm up time 2 A5 H This section provides details about the Modbus protocol Only a subset of the protocol has been implemented as show...

Page 266: ...LSB xxx crc16 lo xxx crc16 hi xxx Example 2 Host Wants to Receive the Batch Setpoint Set point 100 0 Tons Address of Slave Micro Tech 01 Host PLC Slave Micro Tech Address 01H Function 03H Add HI 00H A...

Page 267: ...n 10H add hi xxx add lo xxx n data hi xxx n data lo xxx n bytes xxx data MSB xxx xxx data LSB xxx crc16 lo xxx crc16 hi xxx Response Message address xxx function 10H add hi xxx add lo xxx n data hi xx...

Page 268: ...ave Micro Tech 10H 00H 6CH 00H 02H 04H 00H 00H 42H C8H C4H E4H 01H 10H 00H 6CH 00H 02H 81H D5H Example 5 Message of Sending only one word from Host PC or PLC to Micro Tech code 06H Query address xxx f...

Page 269: ...sage address xxx function 06H add hi xxx add lo xxx value hi xxx value lo xxx crc16 lo xxx crc16 hi xxx Example 6 Host Wants to Write the Batch Number Batch number 0 Address of Micro Tech Slave 1 Host...

Page 270: ...le 1 120 238 Free Space Scale 1 121 240 Free Space Scale 1 122 242 Free Space Scale 1 123 244 Free Space Scale 1 124 246 Free Space Scale 1 125 248 Free Space Scale 1 126 250 Free Space Scale 1 127 25...

Page 271: ...ce Scale 1 202 402 Free Space Scale 1 204 406 Free Space Scale 1 206 410 Free Space Scale 1 208 414 Free Space Scale 1 210 418 Free Space Scale 1 212 422 Free Space Scale 1 214 426 Status Scale 2 216...

Page 272: ...4 566 High Load Set Scale 2 286 570 Low Load Set Scale 2 288 574 Cal Error Scale 2 290 578 Master Total Double Scale 2 Double Float Information 292 582 Reset Total Double Scale 2 Double Float Informat...

Page 273: ...7H Here are some examples showing how the protocol is used Example 1 Connect Message This message must be sent from the Host to Micro Tech Slave before a request of sending data can be performed Host...

Page 274: ...3H BCC DL 10H ACK 06H Example 2 The Host PLC Asks for Data from the Micro Tech This message is sent from the Host to the Micro Tech to request a certain number of data Host PLC Slave Micro Tech DLE 10...

Page 275: ...o download registers to the Micro Tech Host PLC Slave Micro Tech DLE 10H SOH ADDRESS OF SLAVE DLE STX ADDRESS SLAVE ADDRESS HOST COMMAND 0FH STS 00H TMS xxH TMS xxH FNC 67H PACKET OFFSET LOW PACKET OF...

Page 276: ...ro Tech if Data Were Received OK Host PLC Slave Micro Tech DLE 10H ENQ 05H ADDRESS OF SLAVE BCC DLE 10H STX 02H ADDRESS OF HOST ADDRESS OF SLAVE 4FH COMMAND STS 00H TMS xxH TMS xxH DLE 10H EYX 03H BCC...

Page 277: ...C 43 Example 5 The Host Asks the Micro Tech to Send Back the Batch Setpoint Setpoint 200 5 Address of Slave 01 Address of Host 02 Host PLC Slave Micro Tech 10H 01H 01H 10H 02H 01H 02H 0FH 00H 00H 00H...

Page 278: ...ech 9101 9201 Reference Manual Rev J Thermo Fisher Scientific Host PLC Slave Micro Tech 10H 03H 6DH 10H 06H 10H 05H 01H FFH 10H 02H 02H 01H 4FH 00H 00H 00H 99H 09H 06H 94H 08H 00H 80H 48H 04H 10H 03H...

Page 279: ...45 Example 6 The Host Sends the Batch Setpoint to the Micro Tech Batch setpoint 200 5 Tons Address of Slave 01 Address of Host 02 Host PLC Slave Micro Tech 10H 01H 01H 10H 02H 01H 02H 0FH 00H 00H 00H...

Page 280: ...03H 20H 10H 06H 10H 05H 01H FFH 10H 02H 02H 01H 4FH 00H 00H 00H 10H 03H AEH 10H 06H The table below provides more details about the Allen Bradley DF1 registers for Model 9101 9201 Register Address Not...

Page 281: ...N7 12 display 13 N7 13 display 14 N7 14 display 15 N7 15 display 16 N7 16 display 17 N7 17 display 18 N7 18 display 19 N7 19 display 20 N7 20 display 21 N7 21 display 22 N7 22 display 23 N7 23 display...

Page 282: ...atus 3 N7 46 i_o 1 N7 47 i_o 2 N7 48 i_o 3 N7 49 i_o 4 N7 50 i_o 5 N7 51 i_o 6 N7 52 i_o 7 N7 53 i_o 8 N7 54 i_o 9 N7 55 i_o 10 N7 56 i_o 11 N7 57 i_o 12 N7 58 commands N7 59 batch_number N7 60 Load O...

Page 283: ...o Fisher Scientific Micro Tech 9101 9201 Reference Manual Rev J C 49 Register Address Note low_rate_set F8 14 high_speed_set F8 15 low_speed_set F8 16 high_load_set F8 17 low_load_set F8 18 cal_error...

Page 284: ...fused as a simple combination of Ethernet and the Internet Protocol Instead it is an industrial application layer protocol used for communication between industrial control systems PLCs and their comp...

Page 285: ...ple To enter the IP address in the Micro Tech do the following 1 Press the Menu key until Main Menu 4 appears in the display 2 Press the Ether button 3 Press the down arrow button The following defaul...

Page 286: ...AL Operator_Total REAL Batch_Total REAL RX SINT 200 Batch_Setpoint REAL Batch_Preset REAL Batch_Preact REAL High_Rate_Set REAL Low_Rate_Set REAL High_Speed_Set REAL Low_Speed_Set REAL High_Load_Set RE...

Page 287: ...Thermo Fisher Scientific Micro Tech 9101 9201 Reference Manual Rev J C 53 The ladder diagram for the PLC is shown on the following page Ladder Diagram Ladder Diagram Example...

Page 288: ...cientific An example of the Ethernet Module Properties information is shown below Below is an example of the information in the General Tab The information for your specific model Micro tech will be d...

Page 289: ...h 9101 9201 Reference Manual Rev J C 55 On the Connection Tab shown in the graphic below the Requested Packet Interval RPI must be set to 100ms minimum NOTE The Requested Packet Interval RPI must be s...

Page 290: ...Set Success_Flag Both 0 0 Status LED BOTH 1 2 Status Scale 1 2 4 Alarms BOTH 6 12 I O BOTH 12 24 Number Of trials BOTH 24 48 Model ID BOTH 25 50 Batch Number Scale 1 26 52 Batch Deviation Scale 1 27 5...

Page 291: ...Rate Set Scale 1 82 164 Low Rate Set Scale 1 84 168 High Speed Set Scale 1 86 172 Low Speed Set Scale 1 88 176 High Load Set Scale 1 90 180 Low Load Set Scale 1 92 184 Cal Error Scale 1 94 188 Free S...

Page 292: ...tch Total Scale 2 Single Float Information 158 316 Batch Setpoint Scale 2 160 320 Batch Preset Scale 2 162 324 Batch Preact Scale 2 164 328 AZT Correction Scale 2 166 332 Auto Zero Max Dev Scale 2 168...

Page 293: ...8 Batch Deviation Scale 1 5 10 Free Space Scale 1 6 12 Free Space Scale 1 7 14 Free Space Scale 1 8 16 Free Space Scale 1 9 18 Free Space Scale 1 10 20 Free Space Scale 1 11 22 Free Space Scale 1 12 2...

Page 294: ...pace Scale 2 74 148 Free Space Scale 2 75 150 Free Space Scale 2 76 152 Free Space Scale 2 77 154 Free Space Scale 2 78 156 Free Space Scale 2 79 158 Scale Capacity Scale 2 80 160 Speed Capacity Scale...

Page 295: ...Alarms Batch Total FREE SPACE Low Rate Set 11 Alarms Batch Setpoint FREE SPACE High Speed Set 12 I O Batch Preset FREE SPACE Low Speed Set 13 I O Batch Preact FREE SPACE High Load Set 14 I O AZT Corre...

Page 296: ...us Load Cell Capacity Batch Deviation Batch Setpoint 4 Batch Number Rate FREE SPACE Batch Preset 5 Batch Deviation Load FREE SPACE Batch Preact 6 AZ Deviation Speed FREE SPACE Auto Zero Max Dev 7 AZ D...

Page 297: ...FREE SPACE NOT USED NOT USED 28 NOT USED FREE SPACE NOT USED NOT USED 29 NOT USED FREE SPACE NOT USED NOT USED 30 NOT USED FREE SPACE NOT USED NOT USED 31 NOT USED FREE SPACE NOT USED NOT USED 32 NOT...

Page 298: ...does so A Modbus command contains the Modbus address of the device it is intended for Only the intended device will act on the command even though other devices might receive it an exception is specif...

Page 299: ...interface between the master and slaves is structured in parameterization configuration and data transfer phase In the parameterization and configuration phases each slave compares its real configurat...

Page 300: ...evant to wiring and available auxiliary devices like a repeater The table below indicates the maximum cable segment length admitted according to the communication speed Table C 2 Maximum Cable Length...

Page 301: ...entical NOTE The same wires green or red must always be connected to the same terminal A or B in all bus terminals and with all bus connections and be uniform throughout the segment After installing t...

Page 302: ...iguration change Set up data must now be entered Through the new PROFIB menu which can be found on Main Menu 5 it is possible to assign the slave address The limits of values that can be set are the f...

Page 303: ...5 Max 48 Password Service The operator can define how to receive the data from the Micro tech Raw not damped damped or in the same way that the data is displayed Default not damped Selection not damp...

Page 304: ...with a period of 100 milliseconds Errors in the communication are managed in the Micro Tech A specific alarm is generated whenever one or both of the following conditions become true The SPC3 control...

Page 305: ...ters have been required the last time and it will continue to update the read buffer with their actual values What the Master receives after a read operation is a buffer of data having the following s...

Page 306: ...4 5 Data 1 May vary 5 Value of the first register n Data n May vary Value of the last register 1 BLOCK IDENTIFIER This register identifies which block of data the master wants to write The Micro Tech...

Page 307: ...a only if the stamp is different from the stamp received in the previous telegram The master has only to change the stamp value when it creates a new telegram The stamp of the last interpreted telegra...

Page 308: ...0 0 0 0 0 0 1 0 Ext Diag Header 7 0 0 0 0 GE RE LE CE Ext Diag Data GE Group Error Invalid group identifier the group does not exists RE Register number Error The number of requested register is wron...

Page 309: ...diagnostic area Ext_Diag_Data if the bit is set to one If the bit is set to zero a status message can exist in the slave specific diagnostic area Ext_Diag_Data The meaning of this status message depe...

Page 310: ...e request Table C 8 READ BLOCK 0 Status Alarms I O Dynamic data Register Name Offset Start Status LED 5 Status 6 Alarms 10 I O 16 Model ID 28 Zero Number 29 Span Number 31 Rate 33 Load 35 Speed 37 Mas...

Page 311: ...acity 11 Load Cell Capacity 13 AZT Correction 15 Auto Zero Max Dev 17 Total AZ 19 AZT step limit 21 Belt Warm 23 High Rate Set 25 Low Rate Set 27 High Speed Set 29 Low Speed Set 31 High Load Set 33 Lo...

Page 312: ...iation 6 Batch Set 7 Batch Preset 9 Batch Preact 11 Block 102 is the Write block used to manage the registers relevant to Sets and Thresholds Table C 13 WRITE BLOCK 102 Sets and Thresholds Register Na...

Page 313: ...in Long tons per minute LTons The long ton equivalent to 2 240 lbs LTph Long tons per hour Mixed units A menu choice that allows the Micro Tech to display a mixture of English and metric units mV V Mi...

Page 314: ...cro Tech 9201 D07392B E034 Analog I O Board B07392B E003 8 In 8 Out Digital Board B07392B E005 Serial Communication C07392B E008 Siemens Profibus Board C07392B E011 Speed Sensors C07392B E016 Communic...

Page 315: ......

Page 316: ......

Page 317: ......

Page 318: ......

Page 319: ......

Page 320: ......

Page 321: ......

Page 322: ......

Page 323: ......

Page 324: ......

Page 325: ......

Page 326: ......

Page 327: ......

Page 328: ......

Page 329: ......

Reviews: