Initializing the Micro-Tech
Initializing the Software
Thermo Fisher Scientific
Micro-Tech 9101/9201 Reference Manual, Rev J
2-35
To start entering the data the Micro-Tech needs to establish the test
duration for your scale, do the following.
1.
Use the arrow buttons to navigate to the Test Duration screen, which
looks like this.
Nr. = Number.
The zero test can be either “long” or “short.” You should
always
use the
long test when initializing the Micro-Tech. The long and short tests are
assigned the following code numbers.
Long-duration test = 1
Short-duration test = 2
2.
To accept the long-duration test (code = 1), press the down-arrow
button and the following screen appears.
Establish Test
Duration
BATCH
- CAL DATA SCROLL 11 -
Nr. of test duration
> 1 <
READY
ALARM
CALIB
CHOICE
BATCH
- CAL DATA SCROLL 12 -
Establish test duration
READY
ALARM
CALIB
ACQ
MANUAL