Introduction
Hardware Installation
Thermo Fisher Scientific
Micro-Tech 9101/9201 Reference Manual, Rev J
1-11
Provide a cut-out (see
for dimensions) in the panel and, after
removing the holding brackets, and installing the gasket, insert the
Micro-Tech. From the back, insert the holding brackets on both sides of
the Micro-Tech. Tighten the holding brackets to support the Micro-Tech
and form a dust seal.
To connect the incoming power to the Micro-Tech, use the following
procedure. Please note that all units are 24VDC
only
.
1.
For input power, use 16 AWG / 1.5 mmsq standard wires.
2.
Wire the safety ground to the terminal labeled “E” on the Power
Input Terminal.
3.
Wire the +24VDC to the terminal labeled “+” on the Power Input
Terminal.
4.
Wire the 24VDC Common to the terminal labeled “–” on the Power
Input Terminal.
In most instances, your Micro-Tech is shipped to you from the factory
with all the needed jumpers installed and the switches set in the correct
positions for your particular installation and application. As a result, you
should not need to connect any jumpers or set any switches but, if you
do, all the appropriate settings are shown in Appendix A.
NOTE.
Setting jumpers and switches on the Micro-Tech motherboard
must be performed
only
by qualified service personnel.
Mounting
Connecting the
Incoming Power
Supply
Configuring
Jumpers and
Switches
Summary of Contents for Micro-Tech 9101
Page 1: ...Micro Tech 9101 9201 Integrator Reference Manual REC 4281 Rev J Part Number 127336 English...
Page 82: ......
Page 118: ......
Page 124: ...A 2 Micro Tech 9101 9201 Reference Manual Rev J Thermo Fisher Scientific...
Page 172: ......
Page 234: ......
Page 315: ......
Page 316: ......
Page 317: ......
Page 318: ......
Page 319: ......
Page 320: ......
Page 321: ......
Page 322: ......
Page 323: ......
Page 324: ......
Page 325: ......
Page 326: ......
Page 327: ......
Page 328: ......
Page 329: ......