Maintenance and Troubleshooting
Opto22 Problems
4-14
Micro-Tech 9101/9201 Reference Manual, Rev J
Thermo Fisher Scientific
This topic helps you solve problems with the Opto22 output and input
modules. The digital outputs (J7) can be either AC or DC or a
combination of both. Please note that the proper voltage Opto22 must be
used. AC source voltage requires an AC Opto22 module, and a DC
source requires a DC Opto22 module. The digital outputs are controlled
by the Micro-Tech. The LED indicator on the Opto22 module indicates
what the integrator is forcing—that is, Off (the LED is off) or On (the
LED is on).
Here is the procedure to force a digital output.
1.
From the Run screen, press the Menu button three times to navigate
to the “Main Menu 3” screen.
2.
Press the Test button, then press the down-arrow button until the
Digital Output Test screen appears.
Opto22
Problems
Navigating to the
Digital Output
Test Screen
BATCH
- MAIN MENU 3 -
Press MENU for more
READY
ALARM
CALIB
DIA
TEST
BATCH
- TEST SCROLL 4 -
Dig output test
Output #1 OFF
READY
ALARM
CALIB
EDIT
ON/OFF
Summary of Contents for Micro-Tech 9101
Page 1: ...Micro Tech 9101 9201 Integrator Reference Manual REC 4281 Rev J Part Number 127336 English...
Page 82: ......
Page 118: ......
Page 124: ...A 2 Micro Tech 9101 9201 Reference Manual Rev J Thermo Fisher Scientific...
Page 172: ......
Page 234: ......
Page 315: ......
Page 316: ......
Page 317: ......
Page 318: ......
Page 319: ......
Page 320: ......
Page 321: ......
Page 322: ......
Page 323: ......
Page 324: ......
Page 325: ......
Page 326: ......
Page 327: ......
Page 328: ......
Page 329: ......