Doing a material calibration allows you to check the accuracy of your
particular scale. You can do a material calibration in one of two ways, as
follows.
Run material of known weight across your scale and compare the
result with the actual, known weight.
or
Run some material across your scale, note the weight, then transport
the material to another reference scale for re-weighing.
NOTE.
Perform zero and span calibrations prior to Material
Testing.
Material testing requires sufficient material to run without interruption
continually for the required tonnage and time of 6 minutes and 400
counts or 10 minutes, 1,000 Tons (accuracy dependent). Material should
be weighed pre/post material test on a scale proved to within 0.1%.
We are going to run material that we have verified weighs exactly
1,200.00 tons on an independent reference scale.
1.
In the Run screen, press the Menu button and “Main Menu 1” screen
appears.
Material
Calibration
BATCH
- MAIN MENU 1 -
PRESS MENU for more
READY
ALARM
CALIB
SPAN
CAL
ZERO
CAL
MATL
CAL
Summary of Contents for Micro-Tech 9101
Page 1: ...Micro Tech 9101 9201 Integrator Reference Manual REC 4281 Rev J Part Number 127336 English...
Page 82: ......
Page 118: ......
Page 124: ...A 2 Micro Tech 9101 9201 Reference Manual Rev J Thermo Fisher Scientific...
Page 172: ......
Page 234: ......
Page 315: ......
Page 316: ......
Page 317: ......
Page 318: ......
Page 319: ......
Page 320: ......
Page 321: ......
Page 322: ......
Page 323: ......
Page 324: ......
Page 325: ......
Page 326: ......
Page 327: ......
Page 328: ......
Page 329: ......