B-10
Micro-Tech 9101/9201 Reference Manual, Rev J
Thermo Fisher Scientific
R-Cal Span Calibration is described in “Chapter 3 – Operation” (see
page 3-7). Chain and Weights Span Calibrations follow a similar
procedure to R-Cal and can be performed by following the screen
instructions. In addition, the following instructions should be noted:
1.
Make sure the calibration method is set to the appropriate method
(weights or chains) as described above.
2.
Make sure the conveyor belt is
stopped
before applying the weights
or chains.
CAUTION.
Do not
apply weights or chains until instructed to do so by
the Micro-Tech.
3.
When using the weight calibration procedure, if you selected “Load
WTS” in the “I/O Definition Menu” (see page B-38), the following
screen appears.
With the “Load WTS” option selected, the test weights are
automatically loaded after you press the Start button. Wait for the
test weights to be loaded, before proceeding to the next step.
Span Calibration
Instructions
BATCH
AUTO SPAN Weights
Press START to load
the test weights
READY
ALARM
CALIB
EXIT
START
Summary of Contents for Micro-Tech 9101
Page 1: ...Micro Tech 9101 9201 Integrator Reference Manual REC 4281 Rev J Part Number 127336 English...
Page 82: ......
Page 118: ......
Page 124: ...A 2 Micro Tech 9101 9201 Reference Manual Rev J Thermo Fisher Scientific...
Page 172: ......
Page 234: ......
Page 315: ......
Page 316: ......
Page 317: ......
Page 318: ......
Page 319: ......
Page 320: ......
Page 321: ......
Page 322: ......
Page 323: ......
Page 324: ......
Page 325: ......
Page 326: ......
Page 327: ......
Page 328: ......
Page 329: ......