Initializing the Micro-Tech
Initializing the Software
2-20
Micro-Tech 9101/9201 Reference Manual, Rev J
Thermo Fisher Scientific
4.
Press the down-arrow button and the “A/D Channel” screen appears.
The default value is “One” A/D channel. However, if your system has
two load cells, choose the “Two” A/D channels option. Review the
system-specific wiring diagram to determine the number of load cells on
your scale. (The “Type of scale” scroll is not available on a model 9201
Micro-Tech.)
5.
Press the down-arrow button to bring up the units menu.
BATCH
- SC DATA SCROLL 1 -
Number of scales
1
READY
ALARM
CALIB
EDIT
BATCH
- SC DATA SCROLL 1A -
Type of scale
> One A/D Channel <
READY
ALARM
CALIB
CHOICE
BATCH
- SC DATA SCROLL 1A -
Type of scale
> Two A/D Channel <
READY
ALARM
CALIB
CHOICE