Initializing the Micro-Tech
Initializing the Software
2-40
Micro-Tech 9101/9201 Reference Manual, Rev J
Thermo Fisher Scientific
14.
When the span number has been set, the following screen appears
briefly in the display. Notice that the red “Alarm” LED in the
console goes off and the green “Ready” LED comes on.
If you get an “S1 not calibrated” message, check all the numbers you
entered in your Initialization Data Summary table (
). Then go back to the “Entering the Appropriate Conveyor Data”
section above (see page 2-29) and carefully re-enter all the data into the
Micro-Tech. If the calibration fails again, check the load cell (or cells)
are working and sending signals to the Micro-Tech.
15.
After a brief pause, the following screen appears.
BATCH
S1 calibrated
READY
ALARM
CALIB
BATCH
Press RUN to start
or MENU for scrolls
READY
ALARM
CALIB