Operation
Calibrating the Micro-Tech
3-8
Micro-Tech 9101/9201 Reference Manual, Rev J
Thermo Fisher Scientific
In our example, the display shows that the newly established zero is just
0.01% different from the previous zero, meaning that both zeros are
essentially the same and the scale is performing consistently. However,
as there has been a small amount of drift, we decide to reset the zero to
the newly established zero point.
6.
Press the Yes button to accept the new zero, and the zero-changed
screen appears.
7.
Press the Run soft key in the display to return the Micro-Tech to the
Run mode.
8.
Run several zero calibrations to assess the repeatability of the
readings.