Maintenance and Troubleshooting
Load-Cell Problems
4-4
Micro-Tech 9101/9201 Reference Manual, Rev J
Thermo Fisher Scientific
The troubleshooting procedures detailed below consists of four simple
checks, as follows.
Verify excitation/sense voltages.
Verify proper sense jumper settings.
Verify load cell mV signal at the Micro-Tech is within range.
Verify load cell mV signal at the junction box is within range.
The Micro-Tech has a Diagnostics menu to help you troubleshoot the
load cells. Go to “Main Menu 3—Diagnostics.” Two menus pertain to
load-cell troubleshooting, as follows.
A/D Gross and A/D Net (Diagnostic Scroll 1)
Weight on Load Cell (Diagnostic Scroll 2)
This is a digital representation of the mV signal coming from the
load cell. The Micro-Tech maximum A/D is equal to 103,009 at 30
mV. Weight at the load cell can be calculated using either A/D
counts or mV input. Refer to the figure below.
measured DC mV
*
Total load cell capacity
=
Weight at the load cell
exc. Volts * LC mV/V
3 mV/V
A/D Gross
*
Total load cell capacity
=
Weight at the load cell
103009
2 mV/V
A/D Gross
*
Total load cell capacity
=
Weight at the load cell
68673
1.8 mV/V
A/D Gross
*
Total load cell capacity
=
Weight at the load cell
61806
Summary of Contents for Micro-Tech 9101
Page 1: ...Micro Tech 9101 9201 Integrator Reference Manual REC 4281 Rev J Part Number 127336 English...
Page 82: ......
Page 118: ......
Page 124: ...A 2 Micro Tech 9101 9201 Reference Manual Rev J Thermo Fisher Scientific...
Page 172: ......
Page 234: ......
Page 315: ......
Page 316: ......
Page 317: ......
Page 318: ......
Page 319: ......
Page 320: ......
Page 321: ......
Page 322: ......
Page 323: ......
Page 324: ......
Page 325: ......
Page 326: ......
Page 327: ......
Page 328: ......
Page 329: ......