Thermo Fisher Scientific
Micro-Tech 9101/9201 Reference Manual, Rev J
B-11
4.
When using the chain calibration procedure, after applying the chain,
the following screen appears.
After starting the belt, make sure the chain placement is proper
before proceeding by pressing the Start button.
5.
Using R-Cal, Chain, or Weights, after the Span Calibration has been
executed the following screen will display if a material calibration
has been done before and the current simulated load method has no
material factor installed.
Ignore this screen and press Exit. Automatic Material Factoring
should
not
be used. If you want to use the manual Material
Factoring procedure, see page A-12.
6.
Using Chains or Weights, if the Error is below the critical threshold
of 0.50% (0.25% for Model 9201 integrator) press No when the
integrator asks if the span should be changed. If the Error is greater
than 0.50% there may be a problem. Press No and see the
troubleshooting section for additional help.
BATCH
AUTO SPAN Chain
Run belt, the
Press START
READY
ALARM
CALIB
EXIT
START
BATCH
AUTO SPAN COMPLETE
Error X.XX%
Unfactored Calcon
READY
ALARM
CALIB
FACTORS
EXIT
REPEAT
Summary of Contents for Micro-Tech 9101
Page 1: ...Micro Tech 9101 9201 Integrator Reference Manual REC 4281 Rev J Part Number 127336 English...
Page 82: ......
Page 118: ......
Page 124: ...A 2 Micro Tech 9101 9201 Reference Manual Rev J Thermo Fisher Scientific...
Page 172: ......
Page 234: ......
Page 315: ......
Page 316: ......
Page 317: ......
Page 318: ......
Page 319: ......
Page 320: ......
Page 321: ......
Page 322: ......
Page 323: ......
Page 324: ......
Page 325: ......
Page 326: ......
Page 327: ......
Page 328: ......
Page 329: ......