B-16
Micro-Tech 9101/9201 Reference Manual, Rev J
Thermo Fisher Scientific
2.
Press the Material Calibration button and the following screen
appears.
3.
Press the down-arrow button and the Start screen appears.
4.
Make sure that the belt is running empty, then press the Start button.
BATCH
- START OF SCROLL -
Use SCROLL keys to
view selections.
READY
ALARM
CALIB
BATCH
MAT’L CALIBRATION
Run belt empty, then
press START
READY
ALARM
CALIB
MENU
START
BATCH
Run quantity of
Material over scale.
READY
ALARM
CALIB
CONTINUE