Thermo Fisher Scientific
Micro-Tech 9101/9201 Reference Manual, Rev J
B-27
No. Parameter
Selection
Description
Enter time for
revolutions to pass
reference
(10 - 16200 sec)
Time required for minimum number of
revolutions (seconds).
13
Auto Zero tracking
no, yes
14
Auto zero tracking
range +/-
4.0% (0-10)
The AZT sequence starts automatically when
the flow rate is below this percentage of the
max scale capacity.
15
Auto zero tracking
step +/-
0.02% (0.01 – 10 )
16
Auto zero tracking
max dev. +/-
4.0% (0-10);
(0-2)*
(0-5)**
*NTEP W&M mode.
**NTEP W&M mode when print AUTO ZERO
is enabled under the PRINT scroll.
Generates an alarm when the change from
the Zero Reference value to the value of zero
with new factor applied exceeds the max
deviation.
17
Auto zero tracking
Warm Time
1.0 (0-160 min)
18
Auto zero tracking
Test duration
1-L, 1-S
Only available when Nr. of test duration is set
at 2.
19
Max. speed capacity
600.0 (1 - 2000 FPM)
20
Auto Zero max dev.
2% (0-2%)
This scroll is only available when NTEP
W&M mode is selected.
Generates an alarm when the change from
the Zero Reference value to the new zero
value exceeds the max deviation.
21
Zero Reference
INVALID (0 - 120000)
The base value for Autozero max deviation
and Autozero Track max deviation. This
value must be re-set to clear a Autozero or
Autozero Track max deviation alarm.
Summary of Contents for Micro-Tech 9101
Page 1: ...Micro Tech 9101 9201 Integrator Reference Manual REC 4281 Rev J Part Number 127336 English...
Page 82: ......
Page 118: ......
Page 124: ...A 2 Micro Tech 9101 9201 Reference Manual Rev J Thermo Fisher Scientific...
Page 172: ......
Page 234: ......
Page 315: ......
Page 316: ......
Page 317: ......
Page 318: ......
Page 319: ......
Page 320: ......
Page 321: ......
Page 322: ......
Page 323: ......
Page 324: ......
Page 325: ......
Page 326: ......
Page 327: ......
Page 328: ......
Page 329: ......