Maintenance and Troubleshooting
Opto22 Problems
4-18
Micro-Tech 9101/9201 Reference Manual, Rev J
Thermo Fisher Scientific
3.
View Opto LED. With input voltage supplied, LED should be lit.
If GOOD—Replace either the Opto22 input module or the
Micro-Tech CPU board.
If BAD—Replace the Opto22 input module.
Figure 4–11.
Opto22 Troubleshooting—Input Module