Initializing the Micro-Tech
Initializing the Software
Thermo Fisher Scientific
Micro-Tech 9101/9201 Reference Manual, Rev J
2-41
16.
Press the Run button and the Micro-Tech Run screen appears,
which looks like this.
The next step is to go to the “Calibrating the Micro-Tech” section on
page 3-5 to perform the initial zero and span calibrations for your scale.
This is a
very important
step, because the scale will not give accurate
readings until these calibrations are done.
NOTE.
You
must
perform an initial zero and span calibration before
operating your scale.
The Next Step
BATCH
0.0 Tons
0.0 Tph
READY
ALARM
CALIB
TOTALS