Initializing the Micro-Tech
Acquiring Basic System Data
Thermo Fisher Scientific
Micro-Tech 9101/9201 Reference Manual, Rev J
2-7
Measure the distance between the pivot and the first idler, and enter the
result in
. Please note, there may be subsequent
(that is, additional) idlers.
Figure 2–3.
Pivot-to-First-Idler Distance
If the static-weight option is installed, measure the height from the pivot
to the test weight, and enter the result in
. If the
test weight is below the pivot, the value is negative. If this option is
not
available, leave at the default value.
Figure 2–4.
Pivot-to-Test-Weight Height
If the static-weight option is installed, measure the length from the pivot
to the test weight, and enter the result in
. If this
option is
not
available, leave at the default value.
Pivot-to-First-Idler
Distance
Pivot-to-Test-Weight
Height
Pivot-to-Test-Weight
Length