Chapter 15
Maintenance and Troubleshooting
Maintaining a SSL Injector
272
Figure 15-5. Replace and Clean the Liner (3)
13. Clean the septum support from possible fragments left by the septum and
reinsert it into the injector.
14. Place the septum holder on the top of the septum support (see G of
Figure 15-5).
15. Tighten the injector cap to finger tight.
CAUTION
Do not overtighten the injector cap. You could damage the septum and affect
performance.
Restart the GC Conditions
16. In
CARRIER
,
OVEN
and
INLET
menus set the normal working conditions.
17. Perform an automatic leak check.
F
G