Chapter 10
AS 2000 Autosampler Menu
Autosamplers
191
If AS 2000 has been configured for internal standard use (see autosampler
configuration) an additional line is displayed. See also
Pre wash solv
A, B, C, D,
This parameter specifies the solvent vial used for
washing prior the injection.
Pre wash cycles
0–15 times
This parameter specifies the number of times the syringe
is pre-washed with solvent.
Post wash solv
A, B, C, D,
This parameter specifies the solvent vial used for
washing after the injection.
Post wash cycles
0–15 times
This parameter specifies the number of times the syringe
is post-washed with solvent.
Solvent wash vol
depends on
syringe volume
This parameter specifies the volume of solvent used to
clean the syringe.
Plunger strokes
0–15 times
This parameter specifies the number of plunger strokes
to eliminate bubbles.
Viscosity delay
0–15 times
This parameter specifies how long the plunger will
remain at the top of the stroke (to account for viscous
samples).
Sample draw spd
0 - 100 µls
This parameter specifies how quickly the sample is
drawn from the vial.
Air gap volume
0.0–99.9,
100–500 µl
This parameter specifies the amount of air between the
solvent, sample, and internal standard for a solvent
cleaning injection.
Pre dwell time
0–630 seconds
This parameter specifies how long the needle remains
inserted without injecting (hot needle injection).
Post dwell time
0–630 seconds
This parameter specifies how long the needle remains
inserted after injection.
AUTOSAMPLER
Int Std settings
Table 10-1. AS 2000 Autosampler Menu (Continued)
Parameter
Range
Description
Summary of Contents for FOCUS GC
Page 4: ...iv InstructionManual ...
Page 70: ...Chapter 2 The FOCUS GC User Interface Info Diagnostics 70 Instruction Manual ...
Page 92: ...Chapter 4 Column Oven Oven Parameters 92 Instruction Manual ...
Page 124: ...Chapter 6 Purged Packed Column Injector PPKD PPKD Injector Menu 124 Instruction Manual ...
Page 170: ...Chapter 8 Flame Ionization Detector FID Detector Signal Parameters 170 Instruction Manual ...
Page 186: ...Chapter 9 Thermal Conductivity Detector TCD Detector Signal Parameters 186 Instruction Manual ...
Page 242: ...Chapter 13 SOP number P0473 04 E 02 May 2007 242 Instruction Manual ...
Page 336: ...Chapter 16 LAN Option Set up Set Up 336 Instruction Manual ...
Page 344: ...Glossary 344 Instruction Manual C min 1 instead of C min g L 1 instead of g L ...
Page 350: ...Index 350 Instruction Manual ...