Chapter 12
SOP number: P0407/05/E - 31 May 2006
225
17. Press
HOME
to return Main Menu.
18. Activate your Data System and set the parameters required for the checkout. .
19. In
SIGNAL
menu
select Run auto zero
, then press
ENTER
to perform
the automatic zeroing of the system.
Baseline acquisition and Analysis
NOTE
Refer to the Acceptance Values reported in theTable 12-3 according to the data handling in
use.
20. Perform a blank analysis without injecting and press
START
on the GC to
begin the checkout run.
21. With the GC in Stand-by/Prep Run condition, activate the data system for 10
minutes to evaluate your baseline in isothermal condition.
22. After the baseline evaluation has been completed, set-up the data system to
acquire a single run.
23. Inject the test mixture and press
START
on the GC to begin the checkout run.
The resulting chromatogram should look like the one shown in Figure 12-1.
SIGNAL (FID)
Range=10^(0..3)
0
Summary of Contents for FOCUS GC
Page 4: ...iv InstructionManual ...
Page 70: ...Chapter 2 The FOCUS GC User Interface Info Diagnostics 70 Instruction Manual ...
Page 92: ...Chapter 4 Column Oven Oven Parameters 92 Instruction Manual ...
Page 124: ...Chapter 6 Purged Packed Column Injector PPKD PPKD Injector Menu 124 Instruction Manual ...
Page 170: ...Chapter 8 Flame Ionization Detector FID Detector Signal Parameters 170 Instruction Manual ...
Page 186: ...Chapter 9 Thermal Conductivity Detector TCD Detector Signal Parameters 186 Instruction Manual ...
Page 242: ...Chapter 13 SOP number P0473 04 E 02 May 2007 242 Instruction Manual ...
Page 336: ...Chapter 16 LAN Option Set up Set Up 336 Instruction Manual ...
Page 344: ...Glossary 344 Instruction Manual C min 1 instead of C min g L 1 instead of g L ...
Page 350: ...Index 350 Instruction Manual ...