Chapter 15
Maintenance and Troubleshooting
Maintaining a FID
288
OPERATING SEQUENCE
Clean/Replace the FID Collecting Electrode
Material needed
•
Screwdriver
•
Pliers
•
Forceps or tweezers
•
Ultrasonic cleaning bath
•
Liquid detergent
•
Clean compressed air
•
GC-grade methanol
•
Clean paper towel
•
Collecting electrode
WARNING! This operation must be carried out at low temperature to avoid burns. Therefore,
before beginning the sequence, the detector must be cooled to room temperature.
When handling organics solvents you must take precautions to avoid health
hazards.
1. In
DETECTOR
menu, scroll to
Flame
and set OFF.
2. Scroll to
Temp
and set
OFF.
3. Scroll to
H2
and
Air
and set OFF.