Chapter 7
Column
Installation
140
Connecting a Packed Column to the S/SL Injector
Materials required:
•
liner cap removal tool
•
retaining nut
•
ferrules
•
metal adapter
1. Remove the S/SL Injector Top Components.
With reference to Figure 7-6 proceed as follows:
Figure 7-6. Removing the S/SL Injector Top Components
a.
Unscrew the injector cap.
b. Remove the septum holder with septum. then the septum support.
c.
Remove the liner cap by using the tool provided.
d. Use tweezers to remove the liner with the graphite seal.
1. Injector Cap
2. Septum Holder
3. Septum
4. Septum Support
5. Liner Cap Removing Tool
6. Liner Cap
7. Liner
8. Graphite Seal
1
2
3
4
5
6
7
8