Chapter 15
Guide to Solve Analytical Problems
Maintenance and Troubleshooting
317
Behavior
Characteristics
Cause
Remedy
Tailing
(continued)
Sample peaks
(continued)
Glass wool or inlet liner
causing activity
Replace with fresh silanized
wool and a clean inlet liner.
Inlet temperature too low
Increase the inlet temperature
Poor or obstructed column
connections
Remake the column inlet
connection.
Wrong stationary phase
Replace the column according to
the column manufacturer
literature.
Incorrect column position
in inlet
Reinstall the column.
Initial oven temperature too
high (On Column)
Reduce the initial oven
temperature.
Septum purge flow too low
and/or split/splitless vent
flow too low
Check and adjust the septum
purge and vent flows.
Too large injection size
Reduce the injection size.
Unresolved peaks
Carrier gas flow rate too
high
Reduce the carrier gas flow rate
Column deteriorated
Replace the column
Column temperature too
high
Lower the column oven
temperature
Column too short
Use a longer column
Incorrect column choice
Install a suitable column
Injection technique is not
adequate
Choose a correct injection
technique
Summary of Contents for FOCUS GC
Page 4: ...iv InstructionManual ...
Page 70: ...Chapter 2 The FOCUS GC User Interface Info Diagnostics 70 Instruction Manual ...
Page 92: ...Chapter 4 Column Oven Oven Parameters 92 Instruction Manual ...
Page 124: ...Chapter 6 Purged Packed Column Injector PPKD PPKD Injector Menu 124 Instruction Manual ...
Page 170: ...Chapter 8 Flame Ionization Detector FID Detector Signal Parameters 170 Instruction Manual ...
Page 186: ...Chapter 9 Thermal Conductivity Detector TCD Detector Signal Parameters 186 Instruction Manual ...
Page 242: ...Chapter 13 SOP number P0473 04 E 02 May 2007 242 Instruction Manual ...
Page 336: ...Chapter 16 LAN Option Set up Set Up 336 Instruction Manual ...
Page 344: ...Glossary 344 Instruction Manual C min 1 instead of C min g L 1 instead of g L ...
Page 350: ...Index 350 Instruction Manual ...