Chapter 6
Purged Packed Column Injector (PPKD)
PPKD Injector Menu
122
OPERATING SEQUENCE
Programming the PPKD Injector Packed Mode
Before programming the purged packed column injector, do the following:
•
Verify that a column is correctly installed, the correct liner is in the injector, if
used, and the system is free of leaks.
•
Check the oven temperature and detector temperature.
•
Program the carrier gas flow as described in Chapter 3
.
WARNING! Hydrogen is a potentially dangerous gas.
for safety
information.
Proceed as follows:
Use
INC
and
DEC
keys to increase or decrease the parameter values.
1. In Main Menu move the cursor until Inlet is selected. Press
ENTER
to display
the
INLET
menu.
2. In the Mode line select
Packed
and press
ENTER
.
3. Scroll to
Temp
and enter the appropriate injector temperature.
4. If constant septum purge is required, scroll to
Constant Septum Purge
and set On to activate a constant septum purge. If constant septum purge is not
required, set the Constant Septum Purge Off. Scroll to
Stop Purge For:
and enter the time the purge flow should be interrupted.
Summary of Contents for FOCUS GC
Page 4: ...iv InstructionManual ...
Page 70: ...Chapter 2 The FOCUS GC User Interface Info Diagnostics 70 Instruction Manual ...
Page 92: ...Chapter 4 Column Oven Oven Parameters 92 Instruction Manual ...
Page 124: ...Chapter 6 Purged Packed Column Injector PPKD PPKD Injector Menu 124 Instruction Manual ...
Page 170: ...Chapter 8 Flame Ionization Detector FID Detector Signal Parameters 170 Instruction Manual ...
Page 186: ...Chapter 9 Thermal Conductivity Detector TCD Detector Signal Parameters 186 Instruction Manual ...
Page 242: ...Chapter 13 SOP number P0473 04 E 02 May 2007 242 Instruction Manual ...
Page 336: ...Chapter 16 LAN Option Set up Set Up 336 Instruction Manual ...
Page 344: ...Glossary 344 Instruction Manual C min 1 instead of C min g L 1 instead of g L ...
Page 350: ...Index 350 Instruction Manual ...