Using Hydrogen
About This Manual
xxiii
Using Hydrogen
The use of hydrogen as a carrier gas or as fuel for certain flame detectors requires
the operator’s strict attention and compliance with special precautions due to the
hazards involved.
Hydrogen is a dangerous gas, particularly in an enclosed area when it reaches a
concentration corresponding to its lower explosion level (4% in volume). When
mixed with air it can create an explosive mixture. An explosion hazard could
develop in the GC oven when hydrogen is used as a carrier gas if oven elements
are not perfectly connected to each other, or if the connection materials are worn
out, broken, or otherwise faulty.
Use the following safety precautions when using hydrogen:
•
Ensure that all hydrogen cylinders comply with the safety requirements for
proper use and storage. Hydrogen cylinders and delivery systems must
comply with local regulations.
•
Make sure the gas supply is turned completely off when connecting hydrogen
lines.
•
Perform a bubble test to ensure that the hydrogen lines are leak-tight before
using the
instrument.
Perform this test after the pressure test described in the
Testing for Leaks paragraph in FOCUS GC Site Preparation and Installa-
tion Guide Manual. Repeat this test to eliminate all leaks.
•
Ensure your GC column oven has a Thermo Fisher Scientific hydrogen sen-
sor. The hydrogen sensor continuously monitors the hydrogen level in the GC
column oven.
If your GC oven does not have a hydrogen sensor already installed, contact your
Thermo Fisher Scientific sales representative. To comply with instrument safety
requirements, a Thermo Fisher Scientific CSE or authorized service technician
should install the sensor.
If you plan to use a sensor other than the recommended Thermo Fisher Scientific
sensor, you must verify its ability to perform the functions listed above before
installing it. It must comply with your local safety regulations, or with the IEC
61010 regulations if local regulations do not exist.
Summary of Contents for FOCUS GC
Page 4: ...iv InstructionManual ...
Page 70: ...Chapter 2 The FOCUS GC User Interface Info Diagnostics 70 Instruction Manual ...
Page 92: ...Chapter 4 Column Oven Oven Parameters 92 Instruction Manual ...
Page 124: ...Chapter 6 Purged Packed Column Injector PPKD PPKD Injector Menu 124 Instruction Manual ...
Page 170: ...Chapter 8 Flame Ionization Detector FID Detector Signal Parameters 170 Instruction Manual ...
Page 186: ...Chapter 9 Thermal Conductivity Detector TCD Detector Signal Parameters 186 Instruction Manual ...
Page 242: ...Chapter 13 SOP number P0473 04 E 02 May 2007 242 Instruction Manual ...
Page 336: ...Chapter 16 LAN Option Set up Set Up 336 Instruction Manual ...
Page 344: ...Glossary 344 Instruction Manual C min 1 instead of C min g L 1 instead of g L ...
Page 350: ...Index 350 Instruction Manual ...