Chapter 11
Getting Started
Run Time Events
202
Run Time Events
This option allows you to program events to happen prior to or during a run.
The following tables are showing two
RUN TIME EVENTS
menus, one without
entries and one with one entry.
The event shown in the loaded menu concern the detector, it adjusts the detector
Range to the lowest sensitivity during the PrepRun step.
You can program events like this shown with the
RUN TIME
menu and submenu.
Up to 16 events may be programmed.
OPERATING SEQUENCE
Editing Run Time Events
Creating a Run Time Event
Use the following sequence to enter new run time events.
1. In main Menu move the cursor until
Others
is selected, then press
ENTER
.
Scroll to
Run time Events
, then press
ENTER
to open
RUN TIME EVENTS
menu.
RUN TIME EVENTS
RUN TABLE (1 of 1)
<none>
Add run time event
PreRun Det Azero
Add run time event
Clear run time event
Run Time Events menu without
entries
Run Time Events menu with
one entry
Summary of Contents for FOCUS GC
Page 4: ...iv InstructionManual ...
Page 70: ...Chapter 2 The FOCUS GC User Interface Info Diagnostics 70 Instruction Manual ...
Page 92: ...Chapter 4 Column Oven Oven Parameters 92 Instruction Manual ...
Page 124: ...Chapter 6 Purged Packed Column Injector PPKD PPKD Injector Menu 124 Instruction Manual ...
Page 170: ...Chapter 8 Flame Ionization Detector FID Detector Signal Parameters 170 Instruction Manual ...
Page 186: ...Chapter 9 Thermal Conductivity Detector TCD Detector Signal Parameters 186 Instruction Manual ...
Page 242: ...Chapter 13 SOP number P0473 04 E 02 May 2007 242 Instruction Manual ...
Page 336: ...Chapter 16 LAN Option Set up Set Up 336 Instruction Manual ...
Page 344: ...Glossary 344 Instruction Manual C min 1 instead of C min g L 1 instead of g L ...
Page 350: ...Index 350 Instruction Manual ...