Chapter 15
Guide to Solve Analytical Problems
Maintenance and Troubleshooting
311
Baseline
Behavior
Characteristics
Cause
Remedy
Drifting
Accumulation of stationary
phase
Replace the end section of the
column.
Carrier gas cylinder
pressure too low to allow
control
Replace the carrier gas cylinder
or increase the pressure.
Drifting carrier gas or
combustion gas flows
Check the gas controllers.
Accumulation of impurities
in column
Check impurity levels in the gas
source. Use correct gas purity.
Falling
Carrier gas leak in the
system
Perform a leak test and ensure
the tightness of the connections
on the carrier gas line.
Column is baking out
Allow enough time for the
column to stabilize.
Rising
Accumulation of impurities
in column
Check impurity levels in the gas
source. Use correct gas purity.
Contaminated detector
Check the detector and clean it.
Rising under
temperature
program
operation
Column contaminated
Recondition the column.
High-standing current
Carrier gas flow rate too
high
Reduce the carrier gas flow.
Contaminated column
Recondition the column.
Contaminated gases
Replace gas cylinders or gas
filters.
Summary of Contents for FOCUS GC
Page 4: ...iv InstructionManual ...
Page 70: ...Chapter 2 The FOCUS GC User Interface Info Diagnostics 70 Instruction Manual ...
Page 92: ...Chapter 4 Column Oven Oven Parameters 92 Instruction Manual ...
Page 124: ...Chapter 6 Purged Packed Column Injector PPKD PPKD Injector Menu 124 Instruction Manual ...
Page 170: ...Chapter 8 Flame Ionization Detector FID Detector Signal Parameters 170 Instruction Manual ...
Page 186: ...Chapter 9 Thermal Conductivity Detector TCD Detector Signal Parameters 186 Instruction Manual ...
Page 242: ...Chapter 13 SOP number P0473 04 E 02 May 2007 242 Instruction Manual ...
Page 336: ...Chapter 16 LAN Option Set up Set Up 336 Instruction Manual ...
Page 344: ...Glossary 344 Instruction Manual C min 1 instead of C min g L 1 instead of g L ...
Page 350: ...Index 350 Instruction Manual ...