Chapter 11
Instrument Set-up
Getting Started
199
Set Carrier parameters
6. Check that the carrier gas type is correct for the analysis.
In Main Menu move the cursor until
Carrier
is selected. Press
ENTER
to
open
CARRIER
menu.
7. Select the Carrier Constant Flow or Constant Pressure Mode. Enter the initial
Flow or Pressure.
•
If you selected Constant Flow mode, enter in the Flow line the desired
value. The GC calculates the pressure necessary and adjusts the pressure
as necessary to maintain the constant flow.
•
If you selected Constant Pressure mode, then in the Pressure (kPa) line
enter the desired value.
8. Activate the Gas Saver if required. In the Gas Saver Flow (ml/min) and Gas
Saver Time (min) lines the desired values.
NOTE
When you install a new column, you must perform a column evaluation.
Set Oven Parameters
9. In Main Menu move the cursor until
Oven
is selected. Press
ENTER
to open
OVEN
menu.
10. Set the oven temperature and the Oven program required.
Set S/SL Injector Parameters
11. In Main Menu move the cursor until
Inlet
is selected. Press
ENTER
to open
INLET
menu.
12. Set the required temperature setpoint.
13. Select the Split, Splitless or Surge Splitless operating mode.
14. According to the inlet operating mode selected, a different menu will be
displayed. Set the required parameters accordingly.
Summary of Contents for FOCUS GC
Page 4: ...iv InstructionManual ...
Page 70: ...Chapter 2 The FOCUS GC User Interface Info Diagnostics 70 Instruction Manual ...
Page 92: ...Chapter 4 Column Oven Oven Parameters 92 Instruction Manual ...
Page 124: ...Chapter 6 Purged Packed Column Injector PPKD PPKD Injector Menu 124 Instruction Manual ...
Page 170: ...Chapter 8 Flame Ionization Detector FID Detector Signal Parameters 170 Instruction Manual ...
Page 186: ...Chapter 9 Thermal Conductivity Detector TCD Detector Signal Parameters 186 Instruction Manual ...
Page 242: ...Chapter 13 SOP number P0473 04 E 02 May 2007 242 Instruction Manual ...
Page 336: ...Chapter 16 LAN Option Set up Set Up 336 Instruction Manual ...
Page 344: ...Glossary 344 Instruction Manual C min 1 instead of C min g L 1 instead of g L ...
Page 350: ...Index 350 Instruction Manual ...