Chapter 15
Maintenance and Troubleshooting
TCD Troubleshooting
302
Symptom
Diagnosis
Remedy
Low sensitivity
Leaks on the gas lines
Check the connections of the carrier,
reference and make-up gas are tight.
Leak due to septum wearing
Replace the septum.
Operating conditions of the
detector not properly set
Optimize working parameters of the
detector according to the actual
operating mode (Constant Voltage,
Constant Temperature).
Thermal conductivity of the
carrier gas is too close to the
one of the compound to be
analyzed
Use a different carrier gas.
Contaminated filaments
Remove contamination by baking the
filaments for one hour at a
temperature higher than the boiling
point of the most high-boiling
compound.
Perform baking twice if necessary.
If the symptom does not disappears,
contact your customer support
organization.
Refer to Appendix A,
, for contact
information.
Table 15-5. TCD Troubleshooting Table (Continued)
Summary of Contents for FOCUS GC
Page 4: ...iv InstructionManual ...
Page 70: ...Chapter 2 The FOCUS GC User Interface Info Diagnostics 70 Instruction Manual ...
Page 92: ...Chapter 4 Column Oven Oven Parameters 92 Instruction Manual ...
Page 124: ...Chapter 6 Purged Packed Column Injector PPKD PPKD Injector Menu 124 Instruction Manual ...
Page 170: ...Chapter 8 Flame Ionization Detector FID Detector Signal Parameters 170 Instruction Manual ...
Page 186: ...Chapter 9 Thermal Conductivity Detector TCD Detector Signal Parameters 186 Instruction Manual ...
Page 242: ...Chapter 13 SOP number P0473 04 E 02 May 2007 242 Instruction Manual ...
Page 336: ...Chapter 16 LAN Option Set up Set Up 336 Instruction Manual ...
Page 344: ...Glossary 344 Instruction Manual C min 1 instead of C min g L 1 instead of g L ...
Page 350: ...Index 350 Instruction Manual ...