Chapter 15
Maintaining a PPKD Injector
Maintenance and Troubleshooting
279
OPERATING SEQUENCE
Replace or Clean the PPKD Liner
Material needed
•
Glass liner
•
Ultrasonic bath
•
Methanol/acetone mixture (1:1)
•
Tweezers
WARNING! This operation must be carrier out at low temperature to avoid burns. Therefore,
before beginning the sequence, the injector must be cooled to room temperature.
When handling organics solvents you must take precautions to avoid health
hazards.
Preliminary Operations
1. While in Standby condition, in
OVEN
menu set the oven temperature to
30 °C.
2. In
INLET
menu set the inlet temperature to 0 (0 = Off).
3. When the inlet reaches the room temperature, enter in
CARRIER
menu, scroll
to Pressure and set it to 0 (0 = Off).
OVEN
Temp 30 30
INLET
Temp 25 Off
CARRIER
Pressure 0 Off
Summary of Contents for FOCUS GC
Page 4: ...iv InstructionManual ...
Page 70: ...Chapter 2 The FOCUS GC User Interface Info Diagnostics 70 Instruction Manual ...
Page 92: ...Chapter 4 Column Oven Oven Parameters 92 Instruction Manual ...
Page 124: ...Chapter 6 Purged Packed Column Injector PPKD PPKD Injector Menu 124 Instruction Manual ...
Page 170: ...Chapter 8 Flame Ionization Detector FID Detector Signal Parameters 170 Instruction Manual ...
Page 186: ...Chapter 9 Thermal Conductivity Detector TCD Detector Signal Parameters 186 Instruction Manual ...
Page 242: ...Chapter 13 SOP number P0473 04 E 02 May 2007 242 Instruction Manual ...
Page 336: ...Chapter 16 LAN Option Set up Set Up 336 Instruction Manual ...
Page 344: ...Glossary 344 Instruction Manual C min 1 instead of C min g L 1 instead of g L ...
Page 350: ...Index 350 Instruction Manual ...