Chapter 7
Column
Introduction
126
Introduction
The analytical column should be positioned inside the oven. The column ends
should align correctly with the injector and detector base body.
Refer to:
•
Capillary and Wide-bore Columns
•
Capillary and Wide-bore Columns
The capillary column should be positioned inside the oven on the column support.
Using Correct Fittings
To connect a capillary column to the injector and detector base body, you must
use the proper column ferrules and retaining nuts.
Column Ferrules
Graphite ferrules are used for many column connections.
•
Encapsulated graphite ferrules connect the capillary column to the detector
base body and to the injector.
CAUTION
Overtightening compression ferrules does not necessarily produce a stronger,
leak-free joint. In fact, very often the reverse is true. Too much pressure can cause a
leak in the joint and make it very difficult to reseal that particular joint when
changing columns.
Table 7-1 lists the ferrules to use depending on the pre-column and capillary
column external diameter. Ferrules that are the wrong size cause leaks and
contamination.
Summary of Contents for FOCUS GC
Page 4: ...iv InstructionManual ...
Page 70: ...Chapter 2 The FOCUS GC User Interface Info Diagnostics 70 Instruction Manual ...
Page 92: ...Chapter 4 Column Oven Oven Parameters 92 Instruction Manual ...
Page 124: ...Chapter 6 Purged Packed Column Injector PPKD PPKD Injector Menu 124 Instruction Manual ...
Page 170: ...Chapter 8 Flame Ionization Detector FID Detector Signal Parameters 170 Instruction Manual ...
Page 186: ...Chapter 9 Thermal Conductivity Detector TCD Detector Signal Parameters 186 Instruction Manual ...
Page 242: ...Chapter 13 SOP number P0473 04 E 02 May 2007 242 Instruction Manual ...
Page 336: ...Chapter 16 LAN Option Set up Set Up 336 Instruction Manual ...
Page 344: ...Glossary 344 Instruction Manual C min 1 instead of C min g L 1 instead of g L ...
Page 350: ...Index 350 Instruction Manual ...