Chapter 8
Detector Signal Parameters
Flame Ionization Detector (FID)
169
OPERATING SEQUENCE
Setting the FID Signal Parameters
1. In Main Menu move the cursor until
Signal
is selected. Press
ENTER
to
display the
SIGNAL
menu.
2. Scroll to
Range 10^ (0...3)
and set the electrometer amplifier input
range. The value 0 (10
0
) is the most sensitive.
3. Scroll to
Run
Autozero
and set On.
4. If offset is required, scroll to
Offset
and enter a numeric value or set it On
to recall the last offset from memory.
NOTE
If the Range 10^ is set 2 or 3, the small variation of the output signal is not detected. For
this reason, the, Signal pA, Ign. thresh and Flameout retry parameters will be not
displayed in the
DETECTOR
menu.