Chapter 13
SOP number: P0473/04/E - 02 May 2007
238
1). In case of TCD with the polyimide coated filaments, the temperature limit is 320 °C.
After a few second the
ready
message is displayed on
Fil status
line.
14. Press
HOME
to return Main Menu.
15. In Main Menu move the cursor until
Signal
is selected. Press
ENTER
to
open
SIGNAL
menu.
16. Scroll to
Output, Offset, Run autozero, Gain
and
Neg polarity
.
By using the
INC
and
DEC
keys select the value or the required operating
mode, then press
ENTER
.
2. The values in parentheses are not editable
DETECTOR (TCD)
Cost fil temp.
N
DETECTOR (TCD)
Fil volt (CV)
10
DETECTOR (TCD)
Fil temp limit
350
1
DETECTOR (TCD)
Fil status
ready
FOCUS GC
Signal
SIGNAL (TCD)
Output
(1000)
2
Summary of Contents for FOCUS GC
Page 4: ...iv InstructionManual ...
Page 70: ...Chapter 2 The FOCUS GC User Interface Info Diagnostics 70 Instruction Manual ...
Page 92: ...Chapter 4 Column Oven Oven Parameters 92 Instruction Manual ...
Page 124: ...Chapter 6 Purged Packed Column Injector PPKD PPKD Injector Menu 124 Instruction Manual ...
Page 170: ...Chapter 8 Flame Ionization Detector FID Detector Signal Parameters 170 Instruction Manual ...
Page 186: ...Chapter 9 Thermal Conductivity Detector TCD Detector Signal Parameters 186 Instruction Manual ...
Page 242: ...Chapter 13 SOP number P0473 04 E 02 May 2007 242 Instruction Manual ...
Page 336: ...Chapter 16 LAN Option Set up Set Up 336 Instruction Manual ...
Page 344: ...Glossary 344 Instruction Manual C min 1 instead of C min g L 1 instead of g L ...
Page 350: ...Index 350 Instruction Manual ...