Chapter 13
SOP number: P0473/04/E - 02 May 2007
234
OPERATING PROCEDURE
TCD-S/SL Checkout in Splitless Mode
Before starting, verify the configuration of the carrier and set the checkout
required parameters.
Verify Carrier Configuration
1. In Main Menu move the cursor until
Others
is selected, then press
ENTER
.
Scroll to
Configuration
, then press
ENTER
to open Configuration menu
.
Scroll to
Configure carrier
submenu then press
ENTER
.
Select the carrier gas required by using the
INC
and
DEC
keys.
.
2. Press
HOME
three times to return Main Menu.
Set Checkout Parameters
3. In Main Menu move the cursor until
Carrier
is selected. Press
ENTER
to
open
CARRIER
menu.
FOCUS GC
Others
OTHERS
Configuration
CONFIGURATION
Carrier
CONFIGURE CARRIER
Gas type
Helium
FOCUS GC
Carrier
Summary of Contents for FOCUS GC
Page 4: ...iv InstructionManual ...
Page 70: ...Chapter 2 The FOCUS GC User Interface Info Diagnostics 70 Instruction Manual ...
Page 92: ...Chapter 4 Column Oven Oven Parameters 92 Instruction Manual ...
Page 124: ...Chapter 6 Purged Packed Column Injector PPKD PPKD Injector Menu 124 Instruction Manual ...
Page 170: ...Chapter 8 Flame Ionization Detector FID Detector Signal Parameters 170 Instruction Manual ...
Page 186: ...Chapter 9 Thermal Conductivity Detector TCD Detector Signal Parameters 186 Instruction Manual ...
Page 242: ...Chapter 13 SOP number P0473 04 E 02 May 2007 242 Instruction Manual ...
Page 336: ...Chapter 16 LAN Option Set up Set Up 336 Instruction Manual ...
Page 344: ...Glossary 344 Instruction Manual C min 1 instead of C min g L 1 instead of g L ...
Page 350: ...Index 350 Instruction Manual ...