Chapter 10
Autosamplers
AS 2000 Autosampler Menu
190
AS 2000 Autosampler Menu
This paragraph describes the AS 2000 Autosampler menu.
AUTOSAMPLER
Sample volume 1.0
Sample rinses 10
Rinse volume 5.0
Injection speed 100
Pre wash solv A+B
Pre wash cycles 10
Post wash solv A+B
Post wash cycles 10
Solvent wash vol 5.0
Plunger strokes 10
Viscosity delay 10
Sample draw spd 1
Air gap volume 2
Pre dwell time 10
Post dwell time 10
Table 10-1. AS 2000 Autosampler Menu
Parameter
Range
Description
AUTOSAMPLER
This line is the title bar
Sample volume
0.0–99.9,
100–500 µl
This parameter specifies the amount of sample to be
injected for each run.
Sample rinses
0–15 times
This parameter specifies the number of times the syringe
is pre-washed with sample.
Rinse volume
0.0–99.9,
100–500 µl
This parameter specifies the amount of sample or
solvent pulled for each rinse.
Injection speed
1–100 µl/s,
max
This parameter specifies how fast the plunger descends.
Use
max
for speeds over 100 µl/s.
Summary of Contents for FOCUS GC
Page 4: ...iv InstructionManual ...
Page 70: ...Chapter 2 The FOCUS GC User Interface Info Diagnostics 70 Instruction Manual ...
Page 92: ...Chapter 4 Column Oven Oven Parameters 92 Instruction Manual ...
Page 124: ...Chapter 6 Purged Packed Column Injector PPKD PPKD Injector Menu 124 Instruction Manual ...
Page 170: ...Chapter 8 Flame Ionization Detector FID Detector Signal Parameters 170 Instruction Manual ...
Page 186: ...Chapter 9 Thermal Conductivity Detector TCD Detector Signal Parameters 186 Instruction Manual ...
Page 242: ...Chapter 13 SOP number P0473 04 E 02 May 2007 242 Instruction Manual ...
Page 336: ...Chapter 16 LAN Option Set up Set Up 336 Instruction Manual ...
Page 344: ...Glossary 344 Instruction Manual C min 1 instead of C min g L 1 instead of g L ...
Page 350: ...Index 350 Instruction Manual ...