Chapter 8
FID Menu
Flame Ionization Detector (FID)
165
Flame Out Conditions
When the flame is accidentally extinguished, either permanently because of
exhausted fuel gas supplies or temporarily, the
Flame Out
message is displayed
in the menu and a message is recorded in the Run Log.
If the
Retry
function is turned
On
, the system will attempt to re-ignite the flame
up to three times.
Signal pA
0 - 1000.00 pA
Not editable
This parameter shows the collector current in pico
amperes (standing current level). The displayed value is
also used to indicate the flame status. If the value is very
low (such as 0.3 pA), the flame is off. When the value
displayed is greater than the
Ignition threshol
d,
the flame is on.
Ign. thresh
0 - 999 pA
The FID produces a small signal current when lit. This
parameter defines the flame on condition. The FOCUS
GC uses this value to determine flame status (on or off)
and control automatic re-ignition. If
Flameout retry
is
O
n,
the flame will re-ignite if the signal drops below
this value. This line doesn’t appear at detector signal
range 2 and 3. (Threshold assumes 0 pA).
Flameout
retry
On – Off
This indicates re-ignition status. Set On to program
when the flame re-ignition should be attempted. Refer to
for more information. This line
doesn’t appear at detector signal range 2 and 3.
H2
Air
On - Off
These indicate the hydrogen and air flow supplied to the
detector. You can only turn the flows on and off. Set On
to turn on the gas flows. Set Off to turn the flows off.
Make-up
On - Off
This indicates the make-up gas used with the FID. You
can only turn the flows On and Off. Set On to turn on
the gas flows. Set Off to turn off the flow.
Table 8-1. FID Parameters (Continued)
Parameter
Range
Description
Summary of Contents for FOCUS GC
Page 4: ...iv InstructionManual ...
Page 70: ...Chapter 2 The FOCUS GC User Interface Info Diagnostics 70 Instruction Manual ...
Page 92: ...Chapter 4 Column Oven Oven Parameters 92 Instruction Manual ...
Page 124: ...Chapter 6 Purged Packed Column Injector PPKD PPKD Injector Menu 124 Instruction Manual ...
Page 170: ...Chapter 8 Flame Ionization Detector FID Detector Signal Parameters 170 Instruction Manual ...
Page 186: ...Chapter 9 Thermal Conductivity Detector TCD Detector Signal Parameters 186 Instruction Manual ...
Page 242: ...Chapter 13 SOP number P0473 04 E 02 May 2007 242 Instruction Manual ...
Page 336: ...Chapter 16 LAN Option Set up Set Up 336 Instruction Manual ...
Page 344: ...Glossary 344 Instruction Manual C min 1 instead of C min g L 1 instead of g L ...
Page 350: ...Index 350 Instruction Manual ...