Chapter 11
Instrument Start-Up
Getting Started
209
Performing a Manual Injection
1. Perform the Prep Run. by pressing the
START
button a first time. Depending
on the mode you have programmed, the FOCUS GC will perform the
following operation.
•
When the gas saver function is programmed, Prep Run ends the gas
saver mode and resets the split flow to the flow used during the injection.
•
In Splitless mode, Prep Run closes the split valve and will close the
septum purge valve as programmed.
•
In Surge Splitless mode, Prep Run initiates the surge pressure.
2. When the Ready to Inject LED is lit, insert the syringe into the injector, wait
for approximately 3 seconds, inject the sample rapidly, and rapidly remove
the syringe from the injector (This is the Hot Empty Needle technique).
3. Press
START
button.
The GC will complete the analysis as programmed.
Performing Injection Using an Autosampler
Before you begin an autosampler injection, ensure that you have programmed the
autosampler method in the
AUTOSAMPLER
menu and the autosampler sequence in
SEQUENCE
menu.
1. Perform the Prep Run by pressing the
START
button a first time. Depending
on the mode you have programmed, the FOCUS GC will perform the
following operation.
•
When the gas saver function is programmed, Prep Run ends the gas
saver mode and resets the split flow to the flow used during the injection.
•
In Splitless mode, Prep Run closes the split valve and will close the
septum purge valve as programmed.
•
In Surge Splitless mode, Prep Run initiates the surge pressure.
Summary of Contents for FOCUS GC
Page 4: ...iv InstructionManual ...
Page 70: ...Chapter 2 The FOCUS GC User Interface Info Diagnostics 70 Instruction Manual ...
Page 92: ...Chapter 4 Column Oven Oven Parameters 92 Instruction Manual ...
Page 124: ...Chapter 6 Purged Packed Column Injector PPKD PPKD Injector Menu 124 Instruction Manual ...
Page 170: ...Chapter 8 Flame Ionization Detector FID Detector Signal Parameters 170 Instruction Manual ...
Page 186: ...Chapter 9 Thermal Conductivity Detector TCD Detector Signal Parameters 186 Instruction Manual ...
Page 242: ...Chapter 13 SOP number P0473 04 E 02 May 2007 242 Instruction Manual ...
Page 336: ...Chapter 16 LAN Option Set up Set Up 336 Instruction Manual ...
Page 344: ...Glossary 344 Instruction Manual C min 1 instead of C min g L 1 instead of g L ...
Page 350: ...Index 350 Instruction Manual ...