Chapter 15
Maintenance and Troubleshooting
Maintaining a SSL Injector
270
Figure 15-3. Replace a Broken Liner
c.
With the aid of a pipe cleaner, remove the possible glass fragments from
the vaporization chamber.
d. Reinsert the silver seal and the terminal fitting, and then tighten the nut
that retains the terminal fitting. See F of Figure 15-3.
e.
Insert the analytical column with its ferrule into the bottom of the injector
in its previous position, then tighten the M4 retaining nut to hold the
column in place. See G of Figure 15-3.
Replacing the Liner
8. Holding the new (or cleaned) liner with tweezers, place a graphite seal over
the liner, making sure to leave a distance of about 10 mm between the seal and
the liner end.
9. Using tweezers, insert the liner into the injector and push it gently towards the
bottom fitting. See D of Figure 15-4.
D
E
F
G
Summary of Contents for FOCUS GC
Page 4: ...iv InstructionManual ...
Page 70: ...Chapter 2 The FOCUS GC User Interface Info Diagnostics 70 Instruction Manual ...
Page 92: ...Chapter 4 Column Oven Oven Parameters 92 Instruction Manual ...
Page 124: ...Chapter 6 Purged Packed Column Injector PPKD PPKD Injector Menu 124 Instruction Manual ...
Page 170: ...Chapter 8 Flame Ionization Detector FID Detector Signal Parameters 170 Instruction Manual ...
Page 186: ...Chapter 9 Thermal Conductivity Detector TCD Detector Signal Parameters 186 Instruction Manual ...
Page 242: ...Chapter 13 SOP number P0473 04 E 02 May 2007 242 Instruction Manual ...
Page 336: ...Chapter 16 LAN Option Set up Set Up 336 Instruction Manual ...
Page 344: ...Glossary 344 Instruction Manual C min 1 instead of C min g L 1 instead of g L ...
Page 350: ...Index 350 Instruction Manual ...