Chapter 2
Configuration Menu
The FOCUS GC User Interface
53
Oven Configuration Menu
The FOCUS GC oven provides great flexibility in controlling and programming
temperatures. In the
CONFIGURE OVEN
menu you can set various preparatory
parameters.
CONFIGURE OVEN
Max temperature 300
Equilib. Time 10.00
Auto prep run Off
Auto start Off
Prep-run timeout 10
Table 2-4. Configure Oven Menu
Parameter
Range or Options
Comments
CONFIGURE OVEN
This line is the title bar.
Max temp
50–350 °C
in steps of 1 °C
This parameter defines the maximum allowable oven
temperature setpoint to protect the column from
unintentionally high temperatures. This limit must be set
to the manufacturer's maximum recommended operating
temperature for the column.
Equil time
0.00–999.99 min
This parameter defines the equilibration time to stabilize
the oven temperature after the run.
Auto prep run
On/Off
Set On to turn on automatic Prep Run execution without
pressing
START
key. When this item is set to Off, you
must to press
START
key activate the Prep Run.
Auto Start
On/Off
Provides an automatic Start command to GC.
Prep run
timeout
0.00–99.00 min
Enter the duration of the Prep Run. The injection must
occur within this time or the timeout will set the GC to
Standby condition.
Summary of Contents for FOCUS GC
Page 4: ...iv InstructionManual ...
Page 70: ...Chapter 2 The FOCUS GC User Interface Info Diagnostics 70 Instruction Manual ...
Page 92: ...Chapter 4 Column Oven Oven Parameters 92 Instruction Manual ...
Page 124: ...Chapter 6 Purged Packed Column Injector PPKD PPKD Injector Menu 124 Instruction Manual ...
Page 170: ...Chapter 8 Flame Ionization Detector FID Detector Signal Parameters 170 Instruction Manual ...
Page 186: ...Chapter 9 Thermal Conductivity Detector TCD Detector Signal Parameters 186 Instruction Manual ...
Page 242: ...Chapter 13 SOP number P0473 04 E 02 May 2007 242 Instruction Manual ...
Page 336: ...Chapter 16 LAN Option Set up Set Up 336 Instruction Manual ...
Page 344: ...Glossary 344 Instruction Manual C min 1 instead of C min g L 1 instead of g L ...
Page 350: ...Index 350 Instruction Manual ...